GDM-541/533 User Manual
50
Temperature (GDM-533)
Range
Resolution
Accuracy
-40~1000°C
-40~0°C
0.1°C~1°C
± (1.0%+3°C)
0~300°C
± (1.0%+2°C)
300~1000°C
± (1.0%+3°C)
-40~1832°F
-40~32°F
0.2°F~2°F
± (1.0%+6°F)
32~572°F
± (1.0%+4°F)
572~1832°F
± (1.0%+6°F)
The measured temperature should be less than 230°C/446°F.
DC Current
GDM-541
Range
Resolution
Accuracy
220.00μA
0.01μA
± (0.5%+10)
2200.0μA
0.1μA
22.000mA
1μA
220.00mA
10μA
20.000A
1mA
± (1.2%+50)
GDM-533
Range
Resolution
Accuracy
600.0μA
0.1μA
± (1.0%+2)
6000μA
1μA
60.00mA
10μA
± (1.0%+3)
600.0mA
0.1mA
6.000A
1mA
± (1.2%+5)
20.00A
10mA