SPECIFICATIONS
51
Overload protection:
mA/μA range: F1 Fuse 600mA 1000V Φ6x32mm
A range: F2 Fuse 11A 1000V Φ10x38mm
Open circuit allows least significant digit ≤5 (
GDM-533
) and ≤10
(
GDM-541
).
Accuracy guarantee: 1%~100% of range
AC Current
GDM-541
Range
Resolution
Frequency Response
Accuracy
220μA
0.01μA
40Hz~1kHz
± (0.8%+10)
1kHz~10kHz
± (3%+50)
2200μA
0.1μA
40Hz~1kHz
± (0.8%+10)
1kHz~10kHz
± (3%+50)
22mA
1μA
40Hz~1kHz
± (1.2%+10)
1kHz~10kHz
± (3%+50)
220mA
10μA
40Hz~1kHz
± (1.2%+10)
1kHz~10kHz
± (3%+50)
20A
1mA
40Hz~1kHz
± (1.2%+10)
1kHz~10kHz
± (3%+50)
GDM-533
Range
Resolution
Frequency Response
Accuracy
600.0μA
0.1
μA
40Hz~1kHz
± (1.2%+5)
6000μA
1μA
40Hz~1kHz
60.00mA
10μA
40Hz~1kHz
± (1.5%+5)
600.0mA
0.1mA
40Hz~1kHz
6.000A
1mA
40Hz~1kHz
± (2.0%+5)
20.00A
10mA
40Hz~1kHz