SPECIFICATIONS
53
Frequency/Duty Ratio
GDM-541
Range
Resolution
Accuracy
10Hz~220MHz
0.01Hz~0.01MHz
± (0.01%+5)
0.1%~99.9%
0.1%
± (2%+5)
GDM-533
Range
Resolution
Accuracy
10.00Hz~10.00MHz 0.01Hz~0.01MHz
± (0.1%+4)
0.1%~99.9%
0.1%
± (2%+5)
Frequency input amplitude:
≤100kHz: 200mVrms ≤ input amplitude ≤ 20Vrms
>100kHz~1MHz: 600mVrms ≤ input amplitude ≤ 20Vrms
>1MHz (
GDM-533
): 1Vrms ≤ input amplitude ≤ 20Vrms
>1MHz~40MHz (
GDM-541
): 1Vrms ≤ input amplitude ≤ 20Vrms
>40MHz (
GDM-541
): Not specified
Duty ratio measurement is only applicable to square waves.
1Vpp ≤ input amplitude ≤ 20Vpp
Frequency ≤10kHz, duty ratio: 10.0% ~ 90.0%
Overload protection: 1000V