GDM-541/533 User Manual
56
Battery/Fuse Replacement
Battery: 4×1.5V AAA batteries
Fuse: F1 Fuse 600mA 1000V Φ6x32mm (mA/μA input terminal)
F2 Fuse 11A 1000V Φ10x38mm (A input terminal)
When “
” is displayed, please replace the batteries in time
to
ensure measurement accuracy.
Resistance measurement can be used to check the built-in
600mA and 11A fuses. Operation: Insert the red test lead into
the
or
terminal. Insert the red probe into the mA/μA
input terminal to measure the resistance. If the LCD displays
“OL”, the 600mA fuse is blown. Insert the red probe into the A
input terminal to measure the resistance. If the LCD displays
“OL”, the 11A fuse is blown.
Steps of
Replacement
1.
Turn the function dial to the "OFF" position,
and remove the test leads.
2.
Unscrew and remove the battery cover to
replace the batteries and fuses.