GDM-8245
User
Manual
52
Crest Factor Range
3.0 at full scale.
The burden voltage is the same as the DC current.
When the input exceeds the full scale of the selected range, the display will
indicate over-range: “—OL—“.
FREQUENCY MEASUREMENT AT ACA RANGE
RANGE
FREQUENCY
INPUT LEVEL
(SINE WAVE)
ACCURACY
500uA
10Hz ~ 20kHz
≥
90
μ
A
0.05%+1
5mA
10Hz ~ 20kHz
≥
0.9mA
0.05%+1
50mA
10Hz ~ 20kHz
≥
9mA
0.05%+1
500mA
10Hz ~ 20kHz
≥
90mA
0.05%+1
2A
10Hz ~ 2kHz
≥
1A
0.05%+1
20A
10Hz ~ 2kHz
≥
9A
0.05%+1
AC+DC measurement does not support AC+Hz function.
Resistance
RANGE
RESOLUTION
ACCURACY
500
Ω
0.01
Ω
0.1%+4
5k
Ω
0.1
Ω
0.1%+2
50k
Ω
1
Ω
0.1%+2
500k
Ω
10
Ω
0.1%+2
5M
Ω
100
Ω
0.2%+2
20M
Ω
1k
Ω
0.3%+2
Open-circuit Voltage 3.2 volts maximum on 500
Ω
, 1.3 volts maximum on all
other ranges.
Protection
450V dc or peak ac continuous.
Capacitance
RANGE
RESOLUTION
ACCURACY
5nF *
0.001nF
≥
1nF: 2%+10
<1nF &
≥
0.5nF: 2%+20
50nF
0.01nF
≥
10nF: 2%+10
<10nF &
≥
5nF: 2%+30
500nF 0.1nF
2%+4
5uF
1nF
2%+4
50uF 10nF
2%+4
*5nF range is affected by the impedance of the test lead. For accuracy, please
measure the range directly on the input terminal.
Protection
450V dc or peak ac continuous.