REMOTE CONTROL
159
Web Control Interface
The web control interface is accessible with the optional Ethernet card. The
web control interface allows remote access over LAN using a Java-enabled
web browser.
The web control interface allows a web browser to modify parameter settings,
remotely operate, control and monitor the GDM-8261A with a virtual front
panel that mimics the GDM-8261A front panel interface.
Telnet parameters can also be edited by using the web control interface so that
applets such as HyperTerminal or Telnet can be used to monitor
measurement readings, control settings and run programs utilizing the same
remote control command set used with the RS232 remote control.
Background
Before trying to access the web browser control interface,
please ensure your browser has JavaScript and Netbios
enabled.
1. Connection
1.
Configure the LAN interface
and connect the GDM-8261A
to the LAN.
2.
Enter the IP address of the GDM-8261A in the
address field of the web browser.
3.
If WEB PW (web password) is set to ON, a dialog
box will appear prompting for a password. Key in the
password (default password:123456).
4.
The web control Welcome Page appears.
GDM-8261A Welcome Page