GDM-8261A User Manual
54
Voltage/Frequency/Period and Current
Measurement
INPUT
V
W
COM
0~10A
0~1.2A
Load
Note: DC Current measurements will be displayed
as a negative value as the polarity of the current
leads has been reversed.
Please take into account the resistance of the test
leads and internal resistance of the current
connection as it is in series with the test circuit.
The above measuring configuration is used to
measure the voltage present on the resistance
under test and the current through the resistance
under test when using the DCI/DCV or ACI/ACV
dual measurement function.