SYSTEM/DISPLAY CONFIGURATION
91
2.
Press the Down key and then
the Left key twice to choose
A-GAIN. Press the Down key.
The A-GAIN setting appears.
HOLD
HOLD
1st display
Shows A-Zero setting
3.
Select the setting using the
Up/Down key.
4.
Press the Enter key to confirm
your selection. Press the Exit
key to go back to the default
display.
AUTO
ENTER
/ EXIT
SHIFT
5.
A-Zero and A-Gain have an identical time interval
of 5 seconds. As can be seen in the diagram below,
Auto-Gain correction is performed once after
A-Zero has been performed twice.
Expmple
Mode: DCV
Rate: Slow
Accuracy Speed:
650ms
6.
Mode
Rate
Accuracy Speed Quick Speed
DCV, DCI,
4W/2W
S
650ms
495ms
M
217ms
F
70ms
Mode
Rate
Below four measurement modes
don
’
t support either accuracy or
quick speed.
TC, RTD, S
800ms
M
184ms
Diode, Cont S
140ms
M
80ms