SYSTEM/DISPLAY CONFIGURATION
95
Gate Time Setting
Background
The gate time settings determine the accuracy of the
frequency and period measurements. The gate time
settings are the equivalent to the Fast, Medium and Slow
rate settings.
Setting
10ms, 100ms, 1000ms
Rate Settings
The gate time settings are analogous to the rate settings.
Function
Digits
Rate
Readings/s Gate time
Frequency,
Period
6 ½
Slow
1
1000ms
5 ½
Med.
10
100ms
4 ½
Fast
100
10ms
Panel operation
1.
Press the Shift key followed by
the 2nd (Menu) key. Press the
right key three times. The Hz/P
menu appears.
/ EXIT
SHIFT
2ND
MENU
TRIG
TRIG
TRIG
2.
Press the Down key, the right
key and then the down key. The
gate time settings menu appears.
TRIG
1st display
Shows the gate time setting
3.
Select the gate time using the
Up/Down key.
4.
Press the Enter key to confirm
your selection. Press the Exit
key to go back to the default
display.
AUTO
ENTER
/ EXIT
SHIFT