BASIC MEASUREMENT
25
Basic Measurement Overview
Background
Basic measurement refers to the eight types of
measurements assigned to the upper row keys on the front
panel.
ACV
DCV
ACI
DCI
2/4W
/
Hz/P
TEMP
Measurement
type
ACV
AC Voltage
DCV
DC Voltage
ACI
AC Current
DCI
DC Current
2/4W
2-wire and 4-wire Resistance
Diode/Continuity
Hz/P
Frequency/Period
TEMP
Celsius/Fahrenheit Temperature
Advanced
measurement
Advanced measurement (page 55) mainly refers to the
operation using the result obtained from one or more of the
basic measurements.
Refresh Rate
Background
Refresh rate defines how frequently the GDM-8261A
captures and updates measurement data. A faster refresh
rate yields a lower accuracy and resolution. A slower
refresh rate yields a higher accuracy and resolution.
Consider these tradeoffs when selecting the refresh rate.
For DC measurements, the frequency of the refresh rate
depends on the rate settings (S, M, F) and the ADC speed
settings (Accurate, Quick) (page 92 ).
For AC measurements, the refresh rate (S, M, F) is
directly tied to the AC bandwidth settings (page 86).
For further details, please see the specifications.