The DC current range should be manually selected when
the following conditions are true:
1.
When DCI measurement is used.
2.
When the signals being measured contain both DC
and AC components.
3.
When the amplitude of the AC component in the
measured signal is higher or lower than the dynamic
range of the range being currently selected by
auto-range function.
DCI Current Range
Selection List
DCI Range
ADC Dynamic Range
DC 100uA
max±2mA
DC 1mA
max±2mA
DC 10A
max±40mA
DC 100A
max±200mA
DC 1A
max±1.2A
DC 10A
max±10A
2W/4W Resistance Measurement
Measurement
type
2-wire
Uses the standard V-COM ports.
Recommended for measuring resistances
larger than 1k
Ω
.
4-wire
Compensates the test lead effect using the
4W compensation ports, in addition to the
standard V-COM ports.
Recommended for measuring sensitive
resistances smaller than 1k
Ω
.
1. Activate
resistance
measurement
For 2-wire resistance measurement,
press the 2W/4W key once.
2/4W
For 4-wire resistance measurement,
press the 2W/4W key twice.