ADVANCED MEASUREMENT
57
Refresh Rate
Function
S
M
F
(Readings/s)
Continuity / Diode
100
200
300
DCV/DCI/100
Ω
~
100M
Ω
(Accurate)
5
60
240
DCV/DCI/100
Ω
~
100M
Ω
(Quick)
30
600
2400
ACV/ACI
1.2
(sec/reading)
3.38
30
Frequency/Period
1
10
100
Selection steps
1.
Press the Shift key followed by
the AUTO (RATE) key. The
refresh rate switches to the next.
/ EXIT
SHIFT
AUTO
RATE
2.
The refresh rate indicator shows
the current status.
S
→
M
→
F
→
S
Reading Indicator
Background
The reading indicator next to the 1st display flashes
according to the refresh rate when the captured data is
updated on the display.
V
m
When no data is
captured
When there is no captured data, the reading indicator
flashes once every two seconds (slower than the normal
refresh rate), indicating the DMM is in the waiting mode.
Common Attribute: Manual/Automatic Triggering
Automatic
triggering
(default)
The GDM-8261A triggers according to the refresh rate.
See the previous page for refresh rate setting details.