2
z
Initialization
1. Power on the LCR Meter
2. Selects “MANU” mode. How to change MANU mode? Please refer to user manual showing about the panel description (13) in
chapter 3.
3. PC sends command COMU?^END^M (or
^J^M)
4. Waiting for LCR Meter response.
If LCR Meter response is COMU:ON..^END(or
^J^M
) then go to step 5.
If LCR Meter response is COMU:OFF.^END(or
^J^M)
then check below:
(1) LCR Meter‘s baud rate set 38400(default),so PC have to set the same.
(2) LCR Meter’s RS232 item have to be set on. Refer to user manual about OPTION2 in chapter 9-1.
(3) Please check configuration of cable. Refer to user manual about The Configuration of cable in chapter 9-3.
5. PC sends command COMU:OVER^END^M(or
^J^M)
6. Waiting for LCR Meter response. Response is COMU:OVER^END. LCR Meter’s display “RS232 ONLINE”.
7. Initialization is fished.