15
Ex: When R=1kohm,Q= .0005 (R/Q mode and display mode is VALUE)
1. Primary factor of test result (It doesn’t include primary's unit)
LCR Meter response MAIN:PRIM 1.0000^END
ASCII CODE = 4D 41 49 4E 3A 50 52 49 4D 20 20 31 2E 30 30 30 30 0A (Hex format)
Positive symbol
2. Secondary factor and unit
LCR Meter response MAIN:SECO .0005k ^END
ASCII CODE = 4D 41 49 4E 3A 53 45 43 4F 20 20 2E 30 30 30 35 6B 20 0A (Hex format)
Positive symbol Primary factor’s unit
Ex: When R=-1kohm,Q= -.0005 (R/Q mode and display mode is VALUE)
1. Primary factor of test result (It doesn’t includes primary's unit).
LCR Meter response MAIN:PRIM -1.0000^END
ASCII CODE = 4D 41 49 4E 3A 50 52 49 4D 20 2D 31 2E 30 30 30 30 0A (Hex format)
Negative symbol
2. Secondary factor and unit
LCR Meter response MAIN:SECO -.0005k ^END
ASCII CODE = 4D 41 49 4E 3A 53 45 43 4F 20 2D 2E 30 30 30 35 6B 20 0A (Hex format)
Negative symbol Primary factor’s unit