17
EX: If the impedance of “Device-under-test” is small than the existing measurement range of the LCR Meters, LCR Meters will send
message as below:
LCR Meter response PRIM:OV01 ^END
ASCII CODE = 50 52 49 4D 3A 4F 56 30 31 20 0A (Hex format)
C/R mode add a secondary factor’s unit
Ex: When C=.00001nF,R= OVER kohm (C/R mode and display mode is VALUE)
1. Primary factor of test result (It doesn’t include Primary's unit).
LCR Meter response MAIN:PRIM .00001^END
ASCII CODE = 4D 41 49 4E 3A 50 52 49 4D 20 20 2E 30 30 30 30 31 0A (Hex format)
Positive symbol
2. Secondary factor and unit
LCR Meter response SECO:OVER nFk^END Secondary factor’s unit
ASCII CODE = 53 45 43 4F 3A 4F 56 45 52 20 6E 46 6B 0A (Hex format)
Primary factor’s unit