PEL-3000(H) User Manual
138
RPTSTEP
0001 - 1000
Last step number (0001-1000)
per loop
Time Base
0.025 - 600ms
Sets the step execution time.
Data Edit
Overview
Each step in a fast sequence contains the following
setting parameters:
Setting
Setting Range
Description
Step
0001 - 1000
Selects/displays the current
step in the sequence.
The number of available
steps is dependent on the
number of steps added
using the
Ins. Point[F1]
functions.
A minimum of 3 steps.
Value
The current or resistance setting
for the selected operating mode.
TRIG OUT
ON, OFF
When TRIG OUT is set to ON, a
trigger signal is output from the
TRIG OUT BNC terminal at the
start of the step. See page 185
for details.
Time
amplitude
TRIG OUT = ON
TRIG OUT
Start of step
FILL Overview
The FILL function is used to evenly step up the
current or resistance value settings from a starting
step to a finishing step.
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...