PEL-3000(H) User Manual
176
CP Mode
Proportional Control:
Input power = rated power × (external
resistance/10).
Inverse Control:
Input power = rated power × (1 - external
resistance/10).
Input
Power
External
Resistance
Rated
Power
0
Ω
10k
Ω
Proportional control
Inverse control
Note
The inverse configuration is recommended for safety
reasons. In the event that any of the cables become
accidentaly disconnected, the current/voltage/power
input will drop to the minimum. Under similar
circumstances using proportional control, an
unexpectedly high input would result.
Operation
1.
Turn the power off from the PEL-3000(H) and
from the load.
2.
Connect the external resistance across pins 1
and 3 of the J1 connector.
3.
Turn the power on the PEL-3000(H).
4.
Set the operating mode and range.
See page 53 for CC mode.
See page 55 for CR mode.
See page 57 for CV mode.
See page 58 for CP mode.
5.
Press
Main
>
Configure [F5]
>
Next Menu [F4]
>
External [F3]
.
6.
Set the
Control
to
R
for proportional control or
to
Rinv
for inverse control.
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...