PEL-3000(H) User Manual
178
to 10% of the rating in each the operating mode.
Note
Readjustment is needed when you use a different
operating mode, current range or voltage range.
Turning the Load On using External Control
Description
The load can be turned on and off with an external
switch connected to pins 7 and 12 of the J1
connector.
Pin Inputs
Pin 7 of the J1 connector is internally pulled up to
5V with a 10kΩ resistor when the switch is open.
Thus when the switch is open, pin 7 is logically
high. When the switch is closed, pin 7 is pulled
down to the A COM ground level, making pin 7
logically low.
PEL-3000(H)
Switch
Analog
connector
12
7
+5V
A COM
10k
Ω