PEL-3000(H) User Manual
190
Parallel Capacity, PEL-3211H
Model
No. of
Units
V
I
Total Sink Current
PEL-3111H+
PEL-3211H
Total Power
PEL-3111H +
PEL-3211H
PEL-3111H:
Master
x 1
800V 52.5A N/A
N/A
PEL-3211H:
Slave
Boosters
x 1
800V 105A 157.5A
3150W
x 2
800V 210A 262.5A
5250W
x 3
800V 315A 367.5A
7350W
x 4
800V 420A 472.5A
9450W
Note
The PEL-3211H booster packs do not have a control
panel. They can only be used as slaves with a single
PEL-3111H in parallel.
Connection
Description
The J1 and J2 connectors are used for control
during parallel operation. Up to 5 units can be
used in parallel.
Note
Only the rear panel terminals can be used for parallel
operation, the front panel terminals have a lower
current rating and thus should not be used for parallel
operation.
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...