PEL-3000(H) User Manual
226
CP+CV Mode
Load Current
Load Input
Voltage
CP Value
CV level
Note that when the source voltage is less than the
CV level, no current will flow due to a very high
impedance.
CV Mode
CV Mode
When the unit is set to CV mode it will operate as a
constant voltage load when connected to a
constant current source. This means the unit will
maintain a set voltage level, up to the rated power,
regardless of the input current. When the source
voltage is less than the CV level, no current will
flow due to a very high impedance. This is
illustrated below.
CV Mode
Input Current
Load
Voltage
CV level