PEL-3000(H) User Manual
32
in A/us.
Current
Voltage
time
time
∆I
∆T
E
Min. V
The diagram above shows how changes in current
can affect voltage.
Limiting Load line
inductance
Load line inductance can be reduced in two ways.
1.
Ensure load wires are as short as possible and
twist the positive and negative load wires
together.
2.
Current change can be limited by limiting the
slew rate or response speed when switching in
CR and CC mode.
“Twisted pair” will be shown on any connection
diagram where the load wires should be twisted
together.
DUT
-
+
Electronic
Load
+
-
Twisted
pair
DUT
-
+
Electronic
Load
+
-