OPERATION
75
Von Delay
Description
Von Delay is the amount of time the unit will wait
before turning the load on after the Von Voltage
threshold has been latched. This will prevent
overshoot current from affecting the Von Voltage
threshold.
Operation
1.
Press
Main
>
Configure[F5]
>
Other[F2]
and
set the
Von Delay
time.
Range:
Von Delay: OFF, 2.0-60ms(CC, CV, CP
mode)
OFF, 5.0-60ms (2.0-60ms)(CR
mode)
Note
CR mode can have the delay time set separately from
the other modes (called
Von Delay –CR
when in CR
mode).
Timer Functions
Count Time
Description
When Count Time is set to on, it will count the
elapsed time from when the load was turned on to
when it was turned off.
This function is applicable to manual and
automatic shutdown (such as from protection
functions such as UVP etc.)
The elapsed time will be shown in the display
Measurement area.
Operation
1.
Press
Main
>
Configure[F5]
>
Other[F2]
and
turn the
Count Time
on or off.
Range:
ON, OFF
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...