OPERATION
85
Display
8.653
V
50.006
mA
0.4327
W
CV B Value
80.000
CV A Value
80.000
LOAD
RS232
A Value
Fine
01/Oct/2017
V
V
UVP Alarm
Please Press Enter
To Clear Alarm
UVP
UVP indicator
Alarm message
when UVP is
set to Load Off
Mode
CV
I Range
L87.5mA
V Range
L 80V
Response
Fast
Configure
UVP Ring Time
Description
The UVP Ring Time settings allows the UVP alarm
to keep sounding for a user-set amount of time
after the UVP has been tripped.
The alarm will continue ringing for the set amount
of time even if the voltage rises back above the
UVP level~ unless the alarm is cleared manually.
Operation
1.
Press
Main
>
Configure[F5]
>
Protection[F1]
and set the
UVP Ring Time
.
Range:
UVP Ring Time: OFF, 0-600s