OPERATION
95
Go-NoGo
The Go-NoGo configuration is used to create pass/fail limits on the
voltage or current input. If the voltage/current exceeds the
pass/fail limits, an alarm will be output.
The Go-NoGo configuration can be used with the Program function
to create complex pass/fail tests.
Setting the Go-NoGo Limits
Description
The Go-NoGo setting limits can be set as either
discrete high & low values or as a percentage offset
from a center value.
Operation
1.
Press
Main
>
Configure[F5]
>
Go-NoGo[F3].
2.
Select
Entry Mode
and choose how to set the
pass/fail limits.
Value will allow you to set the limits as discrete
values.
Percent will allow you to set the limits as a
percentage offset from a center value.
3.
If
Entry Mode
was set to
Value
, Set the
High
&
Low
limit values.
High:
Low:
0-rating current/voltage
0-rating current/voltage
4.
If
Entry Mode
was set to
Percent
, Set the
Center
voltage/current and
High, Low
% values.
Center:
High:
Low:
0-rating current/voltage
0-100% of center voltage/current
0-100% of center voltage/current
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...