OPERATION
55
CR Mode
Description
In Constant Resistance Mode, the unit will
maintain a constant resistive load by varying the
current. CR mode uses ohms,
Ω
(resistance) or
siemens, S (conductance) for the setting units. For
more details on CR mode, see the appendix on
page 223.
Warning
If you change the mode or the range when the load is
already on, the load will be turned off automatically.
Operation
1.
Make sure the load is off.
2.
Press
Main
.
3.
Select CR mode with the
Mode[F1]
soft-key.
4.
Select the current range with the
I Range[F2]
soft-key.
Range:
High, Middle, Low
5.
Select the voltage range with the
V Range[F3]
soft-key.
Range:
High, Low
6.
Set the resistance or conductance level
parameters using the scroll wheel and number
pad.
For Static mode, set
CR A Value
and/or
CR B
Value
.
For Dynamic mode, set
Level1
and
Level2
.
The maximum and minimum conductance/
resistance levels depend on the selected current
range.
7.
To add CV mode to CR mode (CR+CV), see
page 60.
8.
Set the remaining basic configuration settings
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...