Making Measurements
LT51 User Manual
44
Distortion (THD, S/N, S/ND, SFR)
Distortion measurements can be selected by pressing either the or
key, followed by the [f1] key.
The measurement can be displayed in numerical format, where just the values of
Level, Frequency and distortion for the input signal are displayed, or spectrum
mode, where more information on the type of distortion can be determined.
Once a distortion function has been selected, pressing the or keys,
toggles between the spectral and numerical format displays.
When a Distortion function is active during spectrum displays, the marker is
automatically set to the highest peak displayed (or for SFR, the peak spurious
signal).
Distortion (Spectrum measurements)
Press the key followed by the [f1] key to display Distortion spectrum
measurements:
Press the [f1] soft-key again to cancel any active distortion measurement
and return to
Level
/ Noise
measurements.
Press the [f2] key to measure Total Harmonic Distortion,
THD.
The ratio of the RMS sum of harmonic components (maximum of five) to the
RMS value of the measured input signal in the current frequency span.