trouble shooting
file: n:\article\cos14310m5-v1_07hpc-en\20130207_cos14310m5-v1.07hpc-en_h-p-cosmos_treadmill_manual.doc
© 2013 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 07.02.2013 printed 04.03.2013 page: 134 of 197
Reset of error messages at devices with UserTerminal
Initiation: Running-Belt is not moving, one of the mode LED
is flashing (manual, profile, cardio, test)
Step
Activity
Keys
Response / Display
[01]
Select user option (OP xx)
a d b
Press all 3 keys
simultaneously for
at least 3 seconds
=
indicates: OP01 flashing
!
and
"
indicate: E.rE SEt (for Error
reset)
[02]
Confirm OP01
c
§
indicates: done
Informs you that you have to delete the oil message
[03]
Delete message
c =
indicates: OP 01 flashing
!
and
"
indicate: E.rE SEt (for error
reset)
[04]
Quit user options
b
Standby mode
manual, profile, cardio or test is flashing
With this procedure only the messages E01-OIL and E02-Service can be reset. Other error messages are caused by
technical defects/malfunctions. These error messages can only be reset in the administrator level. In this case please
contact customer service.
Error messages at devices without UserTerminal
Devices without UserTerminal are equipped with an automatic error message code via acoustic signal (beep) indicating
the device has a malfunction, needs lubrication or regular service. The appropriate beep-code sounds in the moment the
error occurs, and afterwards when switching on the running machine if the error still exists three times in a row. To
interpret the error without a display a beep-code is audible according to the error message and according to the list
below:
o = long beep / x = short beep
0= ooooo / 1= xoooo / 2= xxooo / 3= xxxoo / 4= xxxxo / 5= xxxxx / 6 = oxxxx / 7= ooxxx / 8= oooxx / 9= oooox
Example OIL message E01: Acoustic signal beeps 5 times LONG (code for "0") and 1 time SHORT+ 4 times LONG
(code for "1"). This is repeated 3 times. Error codes are explained in the previous chapter.
Reset of error messages on devices without UserTerminal
The OIL message and SERVICE message are quit with OP01. Other error messages have to be quit by OP01 at the
administrator level. In cases, an external UserTerminal or the software h/p/cosmos para control
®
are needed. For further
procedures see the previous chapter.