accessories and options
file: n:\article\cos14310m5-v1_07hpc-en\20130207_cos14310m5-v1.07hpc-en_h-p-cosmos_treadmill_manual.doc
© 2013 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 07.02.2013 printed 04.03.2013 page: 172 of 197
12.18
Retrofitting emergency stop buttons
Emergency stop buttons with magnet holder
external emergency stop button with magnetic
holder for easy fixation on h/p/cosmos running
machines, for example on handrail,
plug: 3 pin XLR male
[cos100548] incl. 10 m cable, no spiral cable
[cos15933] incl. 5 m spiral cord cable.
Emergency stop extern 2 m mount 60 mm (incl. clamp fixation)
external emergency stop for handrail fixation, only for handrails with rail diameter d = 60mm
plug: 3 pin XLR male
To connect this external emergency stop with the treadmill, ensure that your treadmill has a
rear panel terminal for external emergency stops. If there is no socket for external
emergency stops, order the special rear panel terminal for additional keyboard [cos10109]
[cos15021] extern incl. 2 m spiral cord, mount 60mm (incl. clamp fixation)
Emergency stop extern 10 or 15 m (without fixation)
external emergency stop without fixation on the treadmill,
plug: 3-pin XLR plug male
To connect the external emergency stop with the treadmill, ensure that your treadmill has a
rear panel terminal for external emergency stops. If there is no socket for external
emergency stops, order the special rear panel terminal for additional keyboard [cos10109]
[
cos15294L10m
]
incl. 10 m cable, no spiral cable, please stipulate the serial number of the device
[
cos15294L15m
]
incl.15 m cable, no spiral cable, please stipulate the serial number of the device
12.19
Safety multiple socket
The stable safety multiple socket can be used for the safe
connection of a medical h/p/cosmos treadmill with peripheral
devices such as PCs, printer or monitors.
It meets all requirements of EN60601-1 for multiple sockets
for creating connections to a medical electrical system within
the patients environment, i.e. < leakage current of 0.5 mA for
the complete system.
safety multiple socket [cos100157]
N
Use a dedicated line only. Due to the very high load of treadmills, it is never allowed to connect
more than one treadmill or other consumer devices with high load to the multiple safety socket.
N
The power consumption of the consumer devices additionally connected to the treadmill must
be lower than 300 W in total. Furthermore the additional devices must be connected to the
treadmill via a multiple safety socket.
N
When connecting treadmills to the multiple safety socket, there are no performance-limitations
at low-performance applications (walking, slow running). For high-performance applications
(fast running, sprints, etc.), the use of multiple safety sockets may result in performance
limitations.
N
For high performance applications, a 3-phase powered treadmill is recommended.