operation
file: n:\article\cos14310m5-v1_07hpc-en\20130207_cos14310m5-v1.07hpc-en_h-p-cosmos_treadmill_manual.doc
© 2013 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 07.02.2013 printed 04.03.2013 page: 74 of 197
OP19
Setting of sender for POLAR
W.I.N.D. system
0000 0000 = all senders are accepted (can also be set with UP and
DOWN)
xxxx xxxx = only specific sender with special ID is accepted,
must be set with
d
and
a
9999 9999 = next sender will be accepted, saved and filtered (can
also be set with UP and DOWN)
(further settings in user OP23 and administrator OP16 are necessary)
OP20
RS232 interface protocol: COM 1
Various ECG and ergospirometry
systems support the h/p/cosmos
coscom protocol. Individual
adaptation of protocol on request.
For further information about the
h/p/cosmos coscom protocol see
www.coscom.org
We recommend using the
h/p/cosmos coscom v3 protocol
as it provides a high standard of
safety and functionality and is in
compliance with norm EN 62304.
All other interface protocols (other
than coscom v3) are not validated
by h/p/cosmos and not released
by h/p/cosmos based on EN
62304 and EN 14971 and must
not be used for new installations
and configurations of medical
devices and medical applications
from 21.03.2010.
h/p/cosmos para control
®
and
h/p/cosmos para graphics
®
exclusively work with the
h/p/cosmos coscom protocol.
§
indicates flashing: number of the RS232 interface protocol
$
and
%
reports:
OFF = RS232 not active / no protocol / interface deactivated
1 = h/p/cosmos coscom v1, v2, v3 with Baudrate 9600 bps
(standard setting COM 1 and COM 2)
2 = JAEGER OXYCON (up to 1995); if available use:
h/p/cosmos coscom (= 1). Not to be used after 21.03.2010!
3 = printer protocol (serial printer or converter required)
4 = free
5 = free
6 = free
7 = TM treadmill emulation in km/h
if available use: h/p/cosmos coscom (= 1)
8 = TM treadmill emulation in miles per hour
if available use: h/p/cosmos coscom (= 1)
9 = free
10 = Loop Back Test (special test plug required, available
at
h/p/cosmos)
11 = SunTech Tango blood pressure monitor (signal tunnel/loop only)
12 = Remote Control Hardware Terminal MCU 4
(special hardware required)
13 = free
14 = free
17 = free
18 = free
20 = h/p/cosmos coscom v3 with baud rate 115200 bps
For advanced h/p/cosmos coscom v3 connections with baud rate
115200, please select OP20=20. Please note that the connected
devices/software must be approved for h/p/cosmos coscom v3 with baud
rate 115200 (for example h/p/cosmos para control
®
4.1).
OP21
RS232 interface protocol: COM 2 See descriptions above