gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 69
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
5.
In the gaitway-3D[Service] software, select
“Checkout: RMS noise at all speeds” and
follow the instructions.
Without subject on the gaitway-3D, traces of
10 seconds at 1 kHz with the treadmill speed
set at a speed of 0 to 20 km will be recorded
automatically and will generate a plot and
table reporting RMS noise on Fx, Fy and Fz.
Report the plot with the on-site checkout
summary sheet in the treadmill logbook.
6.
In the gaitway-3D[Service] software, select
“Checkout: Natural frequency” and follow the
instructions.
Records of 10 seconds will be taken at 10
kHz while hitting the unloaded treadmill with
a rubber hammer successively in vertical,
fore-aft and lateral directions. A FFT plot
showing resonant frequency and damping
time will be generated automatically.
Report the plot with the on-site checkout
summary sheet in the treadmill logbook.
WARNING: If a pressure platform is present
under the belt, hit the deck beside the belt to
avoid damaging the pressure sensors.