maintenance and safety inspections
file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc
© 2015 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 30.01.2015 printed 30.01.2015 page: 120 of 216
Procedure (for pluto
®
only)
1.
The running belt has to be in standstill (press Stop).
2.
Fill the syringe with 10ml of the lubricant.
3.
Attach the tube to the syringe.
4.
Insert the tube of the syringe under the running belt to the
centre of the running surface.
5.
Pump the lubricant under the running belt slowly.
6.
Remove the syringe including tube.
7.
Start the running machine at a max. speed of 5 km/h for
about one or two minutes.
8.
Repeat this procedure (1 – 7) two more times.
9.
Let the running machine operate for a few minutes at the
speed of 5 km/h, so that the oil is distributed on the inside of
the running belt. Walk for 2 minutes at very low speed (ca. 2
km/h) while varying your position on the running belt. This
will allow the oil under the belt to distribute evenly.
10.
Check the position of the running belt after lubrication. The
belt should be in a centre position on the rear roller (tail
shaft). If the belt needs to be adjusted, please follow the
separate instructions.
11.
After lubrication, the oil message should be reset to "OP01"
(see “Reset of oil-message” in chapter entitled “Optional
functions”). The device has no sensor and the oil message
will not switch off automatically after lubrication.
This oil type must not be used for running decks suited for cycle and wheelchair applications (e.g. saturn 300/100r);
neither should it be used for oil-tanks with an automatic oil-pump. (See separate chapters.)
Special oil 0.25 l – bottle for sliding plate with manual lubrication [cos000200008001]
Syringe 10 ml for lubrication of sliding plate [cos12181]