operation
file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc
© 2015 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 30.01.2015 printed 30.01.2015 page: 28 of 216
5.3
Switch on procedure
5.3.1
Switch on procedure model range 150/50 LC (pluto)
no.
Illustration
Description
[01]
Plug the power cord of the running machine directly
into the wall socket.
The use of extension cables or multiple plug sockets
is not allowed.
[02]
Release all emergency stop push buttons (mushroom
type) of the running machine.
[03]
Make sure that the device protection switch at the
front of the device (front terminal below the hood) is
switched on.
The indicator light (green) within the switch is
illuminated.
[04]
Switch on the running machine by using the silver
push button at the UserTerminal.
The indicator light within the push button is
illuminated (blue).
If the indicator does not flash up, please check the
power supply, the device protection switch and the
emergency stop.
See chapter 9.1 “Device cannot be powered on”
N
If the running machine is at an angle of elevation > 0 when switched on, it will automatically
move to the position zero (display elevation „INIT“). Please ensure that while lowering the
elevation no harm is done to persons or objects under the running machine.
N
Plug the running machines directly into the wall socket. Each running machine should be
connected to a separate circuit. Mark the socket with the name and serial number of the running
machine.
N
The use of extension cables or multiple plug sockets is not allowed.
N
Electrical devices with mains connections must neither be used in wet and humid areas
(e.g. swimming pools, saunas, etc.) nor in environmental chambers.