operation
file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc
© 2015 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 30.01.2015 printed 30.01.2015 page: 80 of 216
OP23
RS232 interface protocol COM 4
§
indicates flashing: number of the RS232 interface protocol
$
and
%
reports:
OFF = RS232 not active / no protocol / interface deactivated
18 = Chip-card reader PROXOMED (special hardware required)
20 = h/p/cosmos coscom v3 / Baudrate 115200 bps
22 = POLAR W:I:N:D: - System (more adjustments in user option OP19
und administrator option OP16)
23 = Chip-card reader ProMedPlus (special hardware required)#
OP27
Minimum acceleration and
deceleration level
The selected minimum level is
valid for all acceleration and
deceleration processes in all
modes and profiles.
§
blinking, reports the min. acceleration / deceleration for all modes
and profiles (standard: level 1)
Level of settings: 1 ... 5 but not higher than the value in option 28. Due to
safety reasons, the acceleration / deceleration levels 5, 6 and 7 cannot
be selected. Note: The selected acceleration and deceleration level is
NOT valid for control and operation via RS232-interface. In this case the
acceleration and deceleration level is set in option no. 29 resp. set in the
corresponding command of the h/p/cosmos coscom protocol.
OP28
Maximum acceleration and
deceleration level
The selected maximum level is
valid for all acceleration and
deceleration processes in all
modes and profiles
§
blinking, reports the max. acceleration / deceleration for all
modes and profiles (standard: level 4)
Fall prevention system prescribed for acceleration level >4.
The maximum acceleration and deceleration level is NOT valid for
control and operation via V24/RS232 interface. In this case the
acceleration and deceleration level is set in option no. 29 resp. set in the
corresponding command of the h/p/cosmos coscom protocol.
OP29
Standard acceleration and
deceleration level for RS 232
interface
The selected acceleration and deceleration level is valid for control and
operation via RS232 interface. This option is very helpful if the peripheral
equipment (e.g. ECG, ergospirometry, PC) does not offer a menu for
acceleration and deceleration levels.
Fall prevention system prescribed for acceleration level >4.
§
blinking, reports: 1 ... 5, (standard: 1) for the acceleration and
deceleration level for all speed commands via RS 232. The maximum
adjustable value depends on the setting of option 28.
Note: If the peripheral equipment sends an acceleration and deceleration
command via the h/p/cosmos coscom protocol, the selected level in
option number 29 is not valid for these speed commands.
OP40
Locking and unlocking of the
treadmill
OFF = After switching on, the treadmill is completely locked / not
accessible. To unlock the treadmill, press the b, - and START
simultaneously. While locked the display shows "no ACCESS"
ON = treadmill is unlocked / accessible (standard)
OP41
Locking and unlocking the manual
mode
OFF = manual mode is locked / not accessible
ON = manual mode is unlocked / accessible (standard)
OP42
Locking and unlocking the profile
mode
OFF = profile mode is locked / not accessible
1 ... 6 = profile mode is unlocked / accessible up to the selected profile
number standard: 6
Example: selected profile number = 3: The profiles 1-3 can be selected,
the profiles 4 – 6 cannot be selected
OP43
Locking and unlocking the cardio
mode
OFF = cardio mode is locked / not accessible
ON = cardio mode is unlocked (standard)