© 1988 - 2019 h/p/cosmos sports & medical gmbh

Handrail, separable [cos14190, cos14191, cos14192]

Title

Description

Short description

Handrails, separable at 1/3 of the running surface

Illustration

Application

Demount the rear part of the handrails to get a better view of the subject (for motion 

analysis, etc.).

Additional safety information

N/A

Technical data

Diameter: 

60 mm

Separable at approx. 1/3 of the running surface

Additional accessories

N/A

Installation

By h/p/cosmos service personnel, only

Further information

https://www.hpcosmos.com/en/products/individual-products/handrail-longshortrolled-

sep13-200

Summary of Contents for venus 200/75

Page 1: ...his document media design software and intellectual property is owned by h p cosmos sports medical gmbh except where explicitly stated differently Warning All rights reserved Unauthorized copying repr...

Page 2: ...ww hpcosmos com Product family treadmill h p cosmos 200 300 75 125 Models venus 200 75 cos30005 01va05 venus 200 75r cos30005 01va06 saturn 250 75r cos30007 01va06 saturn 250 75 cos30007 01va05 venus...

Page 3: ...ic research Many developments and pioneering work from h p cosmos have influenced not only product design and functionality but also their usage and methodologies Your success with our devices is the...

Page 4: ...uration 12 3 5 Contraindications 13 3 Intended Use sport 14 3 1 Intended use 14 3 2 Intended operator 14 3 3 Intended location 14 3 4 Intended duration 14 3 5 Contraindications 15 4 Safety med 16 4 1...

Page 5: ...nterfere with automatic program 39 8 8 Pause function 40 8 9 Acceleration levels 41 8 10 User Options 42 9 Accessories compatible devices 44 9 1 Creating Systems 44 9 2 Overview of accessories 44 9 3...

Page 6: ...58 13 13 Software Programs 58 13 14 All pole disconnection 58 14 Liability and Warranty 59 15 Expected Lifetime 59 16 Disposal 59 17 Annex I 60 17 1 Installation 60 17 2 Commissioning and instruction...

Page 7: ...ions various see below 79 Heart rate measurement POLAR non coded cos102818 80 Heart rate measurement POLAR W I N D coded cos100106 81 Heart rate measurement POLAR H10 cos101787 01 82 Safety arch for t...

Page 8: ...l warning danger warning or caution statements DIN EN ISO 7010 W001 Warning of obstacles stumbling DIN EN ISO 7010 W007 Warning of electrical voltage DIN EN ISO 7010 W012 Warning of hot surface DIN EN...

Page 9: ...y up ISO7000 0623 Keep dry ISO7000 0626 Centre of gravity ISO7000 0627 Temperature limitations SO7000 0632 Do not stack ISO7000 2402 1 2 Symbols used transport 1 3 Structure Illustration Description s...

Page 10: ...rd cos101119 label caution danger zone cos100958 label clean light barrier and reflector weekly cos100958 label h p cosmos address cos10144 01 Umrandung nur Schnittmuster nicht mitdrucken Laufgurteins...

Page 11: ...running deck enables the elevation Both parameters are manually adjustable on the UserTerminal Furthermore operation is possible via pre and self defined modes Operation is also possible via external...

Page 12: ...ect of a medical application is not necessarily a patient Therefore these instructions for use will use the term subject for patients as well as for athletes under test h p cosmos medical treadmills m...

Page 13: ...on has to be made by the medical doctor before the treadmill is used Left main coronary stenosis Main artery disease Cardiac valve disease of moderate severity Known electrolyte imbalance Arterial hyp...

Page 14: ...stop in the following cases while performing sprints for max endurance tests while training on running surfaces wider than 65 cm for children 14 years for subjects with all kind of disabilities impai...

Page 15: ...of moderate severity Known electrolyte imbalance Arterial hypertonia RR 200 mm Hg syst 110 mm Hg diast Tachyarrhythmia or bradyarrhythmia Hypertrophic cardiomyopathy and other forms of outflow tract...

Page 16: ...immediately and consult a medical doctor Exclude overloading or overstressing of the subject The subject has to be checked by a medical doctor before using the device A defibrillator must be present...

Page 17: ...echnical data Do not use the device without instruction by authorized personnel acc to the instruction protocol Regard safety area behind device of 2 0 m x width of treadmill Operator and subject have...

Page 18: ...ibrillator must be present at any time The intended operator has to be in reach of at least one emergency stop off at any time Obey all information given in these instructions for use Do not use the d...

Page 19: ...ame room with the device Operator and subject have to be aware of automatic load changes during profile cardio and test mode Unmeant trapping hazards Take off ties scarfs or other clothes that may be...

Page 20: ...ety arch or a body weight support device airwalk It is up to the operator to use any other device that prevents the subject from falling and complies with IEC60601 1 The pull cord safety stop is not a...

Page 21: ...tion Ilustration Press lever to release rope Check rope daily Renew the rope and harness every 24 months or sooner if it shows any signs of wear or damage Select the user position on the treadmill usi...

Page 22: ...mbles and falls into fall prevention device Treadmill stops Operator subject hits the emergency off Operator helps subject stand up again Operator helps subject exit the device Subject lost consciousn...

Page 23: ...stop Do only use in case of emergency Do not use as normal stop button Do not stand on or enter the running deck when device is in elevation running belt might slip through due to gravity Operation R...

Page 24: ...ll These risks are reduced by safety information within the IFU It cannot be excluded that unintended or forbidden use might cause further not yet regarded risks and that already regarded risks might...

Page 25: ...l prevention Guide subject onto treadmill Do not enter the device when running belt is rotating Do not step on rear roller Do not stand on or enter the running deck when device is in elevation running...

Page 26: ...cord so that the subject has to maintain the correct position see position of subject and operator 5 1 Further preparation locomotion Description Ilustration Adjustment of handrails Pull lever Push bu...

Page 27: ...o mode or UKK walk test The displays may show service information and error messages as well see trouble shooting adjust units and decimals with OP 12 14 Element Primary function Secondary functions S...

Page 28: ...as standard or lt light devices lt devices have no UserTerminal no display no keyboard Standard device lt device Control via UserTerminal Software see accessories additional keyboard see accessories...

Page 29: ...RS232 interfaces must be covered with dust caps cos102973 for isolation purposes Ilustration Description LAN RJ45 Sience port USB 1 USB 2 quick stop power connection Possible network connection possi...

Page 30: ...emergency off button at the UserTerminal body height obstacles etc the operator must install an additional emergency stop within reach see accessories 3 Training area acc to ISO 20957 1 subject devic...

Page 31: ...ght on all emergency offs are released Switch the running machine on with the on off switch on the control panel light goes on Make sure the space under the treadmill is free from persons body parts o...

Page 32: ...ble Manual mode Select mode Confirm and start mode Select speed Select elevation Terminate application Cardio mode Select mode Confirm and start mode Adjust parameters indicated by LEDs Confirm profil...

Page 33: ...h predefined starting speed default 0 5 km h To predefine starting speed see OP09 in User Options To request body weight for calculation of energy and power see OP16 in User Options Confirm with enter...

Page 34: ...4 0 5 10 15 20 25 0 5 10 15 0 5 10 15 20 elev in v in km h t in min profile 2 5 0 5 10 15 20 25 0 5 10 15 0 5 10 15 20 elev in v in km h t in min profile 3 6 Profiles 1 and 4 profile 1 without elevati...

Page 35: ...1 is displayed Select profile with or Current profile is displayed in elevation display Max speed is displayed in speed display Duration is displayed in time display Confirm with enter The selected pr...

Page 36: ...gnals beeps Furthermore the displays show the next load parameters flashing The cardio mode allows training within pre defined heart rate limits In order to stay within the limits the treadmill adjust...

Page 37: ...am The treadmill adjusts speed and elevation according to following matrices Deviation from lower limit Speed km h Elevation Reaction time s 5 heart beats 0 2 0 1 25 6 15 0 4 0 2 25 16 30 0 6 0 4 25 3...

Page 38: ...mation Select test with or Confirm with enter The running belt starts automatically after countdown Following tests require further settings 01 UKK walk test 02 Graded test 03 Conconi test 10 Ramp pro...

Page 39: ...This sets a new max speed Decelerate to 0 km h to pause the operation see pause function Alter elevation Press up or down Profile test mode Only valid for current step Switch between modes Press ente...

Page 40: ...lt stops The speed display shows PAUS Pressing the start key starts the running belt at the preset starting speed default 0 5 km h All values are maintained Pressing the key starts the running belt at...

Page 41: ...acceleration level Example Pressing 3 times then holding results in an acceleration with acceleration level 3 Acceleration levels 1 4 are freely accessible Acceleration levels 5 7 are locked by the a...

Page 42: ...orts medical gmbh Basic functions Buttons displays Further information Press and stop together for at least 3 s speed display shows OP 01 Select option with or Confirm with enter Adjust option with or...

Page 43: ...tem 0000 0000 all senders accepted xxxx xxxx only specific sender accepted 9999 9999 next sender will be accepted OP 20 RS 232 interface protocol COM 1 OFF 1 20 1 h p cosmos coscom OP 21 RS 232 interf...

Page 44: ...illustrations and detailed descriptions see annex or www hpcosmos com Article number Accessory Purpose Information cos10171 h p cosmos safety arch 75 Fall prevention and safety stop Mandatory cos10172...

Page 45: ...spikes 250 125 Ski spike wheel application Optional cos13582 runningbelt for skisticks spikes 300 75 Ski spike wheel application Optional cos00097010041 runningbelt for skisticks spikes 300 100 Ski sp...

Page 46: ...tion Illustration Disinfection Unplug the device Apply some disinfectant to a cloth Wipe all surfaces the subject might have touched Wipe all surfaces that may have come into contact with body fluids...

Page 47: ...Competence Daily inspection Daily Operator Lubrication When OIL message occurs Operator Adjustment of running belt If due Operator Safety inspection 12 months h p cosmos service personnel only Change...

Page 48: ...ding noises during operation The OIL messages occurs every 1000 km by default The oil interval may be varied depending on environmental conditions and use Unplug the treadmill Move one hand tissue for...

Page 49: ...vel Open the lid of the tank and refill the oil tank with approximately 0 5 litre of the provided original special oil to receive at h p cosmos Close the lid of the oil tank Start the running belt wit...

Page 50: ...t to the right Turn counter clockwise to adjust belt to the left After adjustment observe the running belt at 10 km h for at least 2 min Belt must maintain the position Remove Allen key from screw 11...

Page 51: ...e After 12 months or 5000 km there will be a safety inspection reminder see below The error message will be reset by the h p cosmos service personnel performing the safety inspection 11 7 Spare parts...

Page 52: ...stop In case there is no improvement contact service hpcosmos com Elevation does not work E21 Switch off Wait 10 min in order to cool down Switch on again In case E21 still appears unplug device exclu...

Page 53: ...3 on the device see connection of external devices interfaces No connection via RS232 wrong settings Choose correct interface protocol on device User Options OP 20 or 21 Choose correct interface proto...

Page 54: ...for slow or fast change of elevation fast switch is blue lighted belt direction Switch for forward backward running backward switch is blue lighted Belt alignment adjust it Running belt is moving out...

Page 55: ...cm L 250 x W 100 cm L 250 x W 125 cm L 300 x W 75 cm L 300 x W 100 cm L 300 x W 125 cm Track access height 48 cm pit installation on request Handrail dimensions 6 cm Data may be influenced by accesso...

Page 56: ...nderneath the running belt Operation Data Temperature 10 40 C Humidity 30 95 without condensation Pressure 700 1060 hPa Altitude max 3000 m without pressurization altitudes 1000m can cause minor loss...

Page 57: ...utic device and active diagnostic device 60601 1 Protection against electric shock Class I IEC 60601 1 Protection against harmful ingress of water or particulate matter IP20 IEC 60601 1 Mode of operat...

Page 58: ...t prior notice E OE errors and omissions excepted Note In case of special voltage versions the data on the name plate apply Connect mains only if data on nameplate and local power supply are identical...

Page 59: ...com Limited liabilities apply If h p cosmos or h p cosmos organizational bodies senior management or agents can be held accountable for the payment of damages pertaining to slight negligence breach o...

Page 60: ...E bolt If in compliance with local national standards use type A RCDs residual current operated protective device For type B RCDs measuring also direct current an additional leakage current limiter ha...

Page 61: ...are and connectors are explained in detail UserTerminal 10 The general use is explained incl manual profile cardio and test mode Operation 11 The competences and intervals for maintenance are explaine...

Page 62: ...ty information intended or forbidden use as well as unauthorized maintenance or lack of maintenance and regular safety checks may lead to injury or even death and can damage the device Furthermore thi...

Page 63: ...istance covered in km indicates km OP03 Indication of total hours of operation stand by time including runtime of motor running belt h and show Operation hours reports h OP04 Indication of total hours...

Page 64: ...mation of power and energy consumption OP16 Request for body weight before manual or automatic start 0 OFF Request for body weight before starting a program is not required Calculation of energy consu...

Page 65: ...ps For advanced h p cosmos coscom v3 connections with baud rate 115200 please select OP20 20 Please note that the connected devices software must be approved for h p cosmos coscom v3 with baud rate 11...

Page 66: ...the h p cosmos coscom protocol the selected level in option 29 or in the corresponding command of the h p cosmos coscom protocol OP40 Locking and unlocking of the treadmill OFF After switching on the...

Page 67: ...after pressing START again and will not be automatically deleted by pressing STOP Display values can only be deleted by pressing the STOP key twice time distance energy OP46 Report mode display Eleva...

Page 68: ...erforming the UKK walk test the subject must warm up and determine the max walking speed During the test the subject must walk as fast as possible for 2 km heart rate approx 80 of max The subject must...

Page 69: ...by the medical doctor Skip remaining break time Press start once restart after countdown Press start twice restart immediately Prolong break Press within break pause is indicated Press start to contin...

Page 70: ...ion 1 3 00 2 7 10 2 4 0 12 3 5 4 14 4 6 7 16 5 8 0 18 6 8 8 20 7 9 6 22 Description Illustration e g for ECG stress test Regard chapter test mode for safety information and adjustment Step Duration mi...

Page 71: ...o test mode for safety information and adjustment Step Duration min Speed km h Elevation 1 3 00 2 7 10 0 2 4 8 10 0 3 6 4 10 0 4 8 0 10 0 5 8 0 15 0 6 9 6 15 0 18 7 Cooper protocol Description Illustr...

Page 72: ...t steps onto the running belt the doctor presses the START key again By pressing the START key the second time the displays will be reset to zero Step Duration min sec Speed km h Elevation Total time...

Page 73: ...er for at least 5 s Select speed for step 1 with or Confirm with enter Select distance for step 1 with or Confirm with enter Select distance 0 in order to program this step by time Select time for ste...

Page 74: ...isplays Further information Select acceleration for step 1 with or Confirm with enter The acceleration level is displayed in the index display Proceed the same way for further steps Make sure all step...

Page 75: ...n the PC and keys can be simulated for remote control of the running machine or the ladder ergometer Illustration Application See separate instructions for use Additional safety information See separa...

Page 76: ...Application The additional keyboard allows full control over all functions of the MCU like START STOP elevation UP and DOWN speed and Additional safety information See safety information for the trea...

Page 77: ...hement 10m cos15294 L15m Emergency stop ext without attachement 15m Illustration with magnet holder without attachement Application Operation Result Release Restart Push button Running belt stops with...

Page 78: ...description Increase of maximum speed to 50 60 80 km h Illustration N A Application No changes Additional safety information N A Technical data Speed cos10028 0 50 km h cos00096110029 0 60 km h cos00...

Page 79: ...cation No changes Additional safety information N A Technical data Thickness 5 mm Length x width of running surface cos13587 200 x 75 cm cos13586 200 x 100 cm cos13585 250 x 75 cm cos13584 250 x 100 c...

Page 80: ...han POLAR T31 or T34 is used see print If the heart rate is not displayed or displayed incorrectly There might be interferences with Screens computers printers mobile phones and any radio engineering...

Page 81: ...sing immediately and consult a medical doctor Technical data Transmission radius approx 10m Further data see accompanying POLAR documents or www polar com Troubleshooting In case the heart rate is not...

Page 82: ...zy stop exercising immediately and consult a medical doctor Technical data Further data see accompanying POLAR documents or www polar com Troubleshooting In case the heart rate is not displayed Chest...

Page 83: ...vation 288 cm 2 cm buffer zone Running machines with 250 cm or 300 cm length of running surface 272 cm with 0 elevation 270 cm 2 cm buffer zone 300 cm with maximum elevation 298 cm 2 cm buffer zone Ad...

Page 84: ...Illustration Application Demount the rear part of the handrails to get a better view of the subject for motion analysis etc Additional safety information N A Technical data Diameter 60 mm Separable at...

Page 85: ...onal balance control Illustration Application Subject may hold crossbar front rail for balance control Holding handrails during use affects exercise results Additional safety information It is not all...

Page 86: ...jump on and off the running belt during speed and sprint training Additional safety information Use of safety arch is obligatory No other person may be located on the extended footboard Footboard is o...

Page 87: ...Illustration Application Attach part 1 to treadmill and lock it Attach part 2 to the front part and lock it Additional safety information Using a wheelchair on the treadmill is only allowed on treadm...

Page 88: ...pplications with wheels as well Applications with wheels include cycling roller skiing wheelchair applications etc For any wheelchair application a fall prevention device is prescribed and absolutely...

Page 89: ...formation Only devices marked with an r or rs like h p cosmos saturn 250 75 r are intended for applications with wheels as well Applications with wheels include cycling roller skiing wheelchair applic...

Page 90: ...erman mobile networks fax 49 18 05 16 76 69 email service hpcosmos com Skype hpcosmos com search select name Sales phone 49 18 05 16 76 67 0 14 min from German landlines max 0 42 from German mobile ne...

Reviews: