background image

11

DEUTSCH

ENGLISH 

FRANÇAIS

ITALIANO

ESPAÑOL

NEDERLANDS

PORTUGUÊS

SVENSKA

DEUTSCH

ENGLISH

FRANÇAIS

ITALIANO

ESPAÑOL

NEDERLANDS

PORTUGUÊS

SVENSKA

© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7006000‑04270 – 2022 – 09

6 Operation

WARNING!

Reusable measuring prisms must be cleaned and disinfected after every

examination.

6.1 Astigmatism

• If the cornea is spherical, measurements can be taken on any meridian, but it is

most convenient to measure on the 0° meridian.

• However, the choice of meridian is significant when measuring eyes with corneal

astigmatism greater than 3 dioptres, since the flattened area is not circular, but

elliptical.

• It has been calculated that, in cases of greater corneal astigmatism, a surface of

7.354 mm

2

(ø 3.06 mm) is applanated if the measuring prism is at an angle of

43° (mark (A)) to the meridian of the greatest radius.

Example:

If the corneal astigmatism measures

6.5 mm / 30° = 52.0 dpt / 30° and

8.5 mm / 120° = 40.0 dpt / 120°

the 120° graduation mark on the prism is set on the mark

(A) of the prism holder.
If, on the contrary, the measurements are

6.5 mm / 120° = 40.0 dpt / 120° and

8.5 mm / 30° = 52.0 dpt / 30°

then the 30° graduation mark is set on the mark (A), mean-

ing that simply the axis position of the largest radius is

placed on the mark (A).

6.2 How pressure is measured

The applanation tonometer functions according to the 'Goldmann method': measure-

ment of the pressure required to maintain a uniform applanation of the surface of the

cornea.

• The readings obtained are based on the assumption that a cornea with a 'nor-

mal' thickness is being measured; a different corneal thickness leads to changes

in the measured IOP. A 'normal' corneal thickness is considered to fall within the

range of 530 to 560 micrometre.

• If it is suspected that measuring results may be inaccurate, please conduct a

functional verification in accordance with the 'Maintenance' chapter.

6.3 Preparing the patient

WARNING!

Measurements without fluorescein can lead to false results.

NOTE!

A smooth operation can only be guaranteed with the use of original

Haag-Streit measuring prisms and Tonosafe.

1. Anaesthetise each eye

2. Apply Fluorescein into the eye to be examined

3. The correct eye height of the patient can be set via the chin rest.

6.4 Patient instructions

1. Press the head firmly against the chin and forehead rests.

2. The patient must look straight forward. If necessary, the small fixing light can

be used to steady the eyes.

3. It is recommended that the patient be repeatedly reminded to keep his eyes

open wide during the examination. It might be necessary for the examiner to

keep the eye open by splaying the eyelids with thumb and index finger.

4. This must be done without applying any pressure to the eye.

6.5 Preparation of slit lamp and tonometer

For all Haag-Streit slit lamps and tonometers

1. Before the examination, the eyepieces must be adjusted correctly to the exam-

iner.

2. Set the magnification to 10×.

3. Adjust the illumination to medium intensity.

4. Position the blue filter in the beam path of the slit lamp's illumination apparatus

and open the slit diaphragm fully.

5. Insert the disinfected measuring prism into the (0° position) holder on the

sensor arm. For Tonosafe, see the separate instructions for use.

6. Snap the sensor arm into place so that the axes of the measuring prism and of

the microscope align.

7. Set the measuring drum to the first calibration line.

Summary of Contents for AT 900

Page 1: ...GLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7006000 04270 2022 09 INSTRUCTIONS FOR USE Applanation tonometer AT 900 AT 870 27 Edi...

Page 2: ...ly with the regulations in these instructions for use we can guarantee reliable and trouble free use of our product WARNING Applanation tonometry may only be conducted by appropriately trained and qua...

Page 3: ...ment 8 2 4 Clinical benefit 8 3 Introduction 9 3 1 Overview 9 4 Device assembly installation 9 4 1 AT 900 model R 9 4 2 AT 900 model T 9 4 3 AT 900 model BQ 9 4 4 Tonometer AT 870 9 4 5 Which slit lam...

Page 4: ...ing shows traces of im proper handling or damage If this is the case notify the transport company that delivered the goods to you Unpack the device together with a representative of the transport comp...

Page 5: ...hanges the biomechanical properties of the cornea The biomechanical properties of the cornea have an effect on the measurement readings 1 6 Disinfection NOTE The applanation tonometer can but does not...

Page 6: ...The operator accepts all liability for the use of other disinfection agents Single use prisms must be replaced after every examination see separate instructions for use The above mentioned instruction...

Page 7: ...the product This would void all guarantee claims Continued use of a damaged product may lead to personal injury In such a case the manufacturer will not accept any liability Haag Streit does not gran...

Page 8: ...2 3 1 Contraindications The list of known contraindications includes but is not limited to the following medic al conditions Active corneal infections Corneal epithelial defects Ocular chemical burn...

Page 9: ...trained specialists 4 1 AT 900 model R A pivot is required for use of the Applanation Tonometer AT 900 model R on the BM 900 slit lamp It is assembled as follows 1 Unscrew the mounting screw found at...

Page 10: ...ts from the bottom side into the beam path of the microscope and the illumination The Applanation Tonometer AT 900 model BQ has a working position and two resting positions To set the correct angle of...

Page 11: ...red IOP A normal corneal thickness is considered to fall within the range of 530 to 560 micrometre If it is suspected that measuring results may be inaccurate please conduct a functional verification...

Page 12: ...through observation with the wrong eyepiece adjustment can lead to false results 6 6 Measuring correctly NOTE Make sure that the measuring arm of the tonometer can move freely and is not hindered by t...

Page 13: ...he measuring prism is touching the eye only intermittently If the patient pulls back even further the fluorescein rings will disappear com pletely Instruct the patient to adopt the correct position On...

Page 14: ...mp up wards Nearly entire circle above incomplete circle below 11 Measuring prism not centred on the eye eye still too high Using the control lever move the slit lamp up wards Two incomplete circles t...

Page 15: ...ght Installation AT 900 model R Can be fitted on the peg on the microscope AT 900 model T Can be fitted on the guide plate on the swivelling peg for the microscope and illumination arm AT 900 model BQ...

Page 16: ...evice should be cleaned weekly as described We recommend having the device checked annually by an authorised service tech nician 8 2 Cleaning and disinfection Tonometer housing and sensor arm can if r...

Page 17: ...AT 870 Check the drum position 0 Check position 0 05 The zero line on the measuring drum is moved downwards opposite the index by a single marking width The feeler arm must move by itself all the way...

Page 18: ...suring drum s calibration mark 6 down wards opposite the index by a interval The feeler arm must move towards the examiner Check position 6 1 The check point is at 6 1 Move the measuring drum s calibr...

Page 19: ...on of the patient The feeler arm must move towards the patient Check position 1 95 Move calibration mark 2 on the measuring drum downwards opposite the fixed index by a width of one marking With light...

Page 20: ...n of the measuring prism and the separate instructions for use 9 3 Tonosafe disposable prisms Tonosafe disposable prisms are a comfortable and ef fective solution to reduce the risk of cross infection...

Page 21: ...ould you have any further questions please contact your Haag Streit representative at www haag streit com haag streit group contact haag streit distributors distributors HAAG STREIT AG Gartenstadtstra...

Reviews: