background image

15

© HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. No. 1500.7200492-04150 – 15. Edition / 2017 – 01

DEUTSCH

ENGLISH 

FRANÇAIS

ITALIANO

ESPAÑOL

NEDERLANDS

PORTUGUÊS

SVENSKA

DEUTSCH

ENGLISH

FRANÇAIS

ITALIANO

ESPAÑOL

NEDERLANDS

PORTUGUÊS

SVENSKA

F.5  Safe distances for non-life support devices (standard table 6)

The information is based on the requirements of EN 60601-1-2:2007 (IEC 3

rd

 edition).

Recommended safe distances between portable and mobile HF communication devices and this device.
This product is designed to be operated in an electromagnetic environment in which radiated HF interference is controlled. The customer or user of this product can help to 

prevent electromagnetic interference by maintaining minimum distances between portable and mobile HF communication systems (transmitters) and this product, as rec-

ommended below in accordance with the maximum output of the communication system.

Nominal output of the transmitter (W)

Safe distance according to transmission frequency (m)

150 kHz – 80 MHz

= 0.35 

80 MHz – 800 MHz

= 0.7 

800 MHz – 2.7 GHz

= 1.4 

0.01

0.1

1

10

100

For transmitters with a nominal output not listed in the table above, the distance 

D

 can be calculated in meters (m) using the equation for the respective  column, in which 

P

 

is the nominal output of the transmitter in watts (W)
NOTE 1: At 80 MHz and 800 MHz the higher frequency applies.
NOTE 2: To calculate the recommended safe distance of transmitters in the frequency range of 80 MHz to 2.7 GHz an additional factor of 

10

/

3

 was used to reduce the 

probability of a mobile/portable communication device causing interference if inadvertently brought into the patient area.
These guidelines may not apply in all situations. Electromagnetic wave propagation is influenced by absorption and reflection of buildings, objects and people.

NOTE 3:

0.035

0.1

0.35

1.1
3.5

0.07

0.2
0.7
2.2

7

0.14
0.44

1.4
4.4

14

01_IFU_BM900-7200492-04150_eng.indd   15

21.02.2017   09:55:00

Summary of Contents for BM 900

Page 1: ...00 7200492 04150 15 Edition 2017 01 DEUTSCH ENGLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA DOC no 1500 1500 1400209 04000 INSTRUCTIONS FOR USE Slit lamp BM 900 15 Edition 2017 01 01_IF...

Page 2: ...of diseases or trauma which affect the structural properties of the eye Contraindication There is no absolute contraindication for tests with this device Appropriate profes sional judgement and cautio...

Page 3: ...Preparing the patient 7 5 3 Operating the instrument 7 5 4 Setting the filters diaphragms 8 5 5 Fixation star 9 5 6 Microscope and eyepiece 9 6 Decommissioning 9 7 Technical data 9 7 1 Slit illuminat...

Page 4: ...npack the appliance check whether the packaging shows traces of incorrect handling or damage If this is the case notify the transport company that has delivered the goods to you Unpack the equipment t...

Page 5: ...nued use of a product damaged by incorrect handling may lead to personal injury In such a case the manufacturer will not accept any liability Haag Streit does not grant any warranties either expressed...

Page 6: ...Breath shield option 14 Protective cover 15 Illumination unit microscope an gle scale 16 Illumination arm locking screw 17 Microscope arm locking screw 18 Slit width setting screw 19 Slit length diap...

Page 7: ...s and press the rocker switch The green light on the rocker switch illuminates when the device is switched on Turn the rotating knob on the illumination control to a position between 1 and 10 5 Operat...

Page 8: ...d 10 a short mirror 12 is used the illumination turned 90 using the locking screw 20 and tilted in 5 steps us ing the latch 27 and the illumination and microscope turned to the central po sition 0 The...

Page 9: ...Dioptre scale for 16x eyepieces for setting the refraction of the examin er 8 D 46 Dioptre scale for 10x eyepieces for setting the refraction of the examin er 8 D 47 Index white point 48 10x or 25x e...

Page 10: ...from the to the side The structure of the test rod must be in sharp focus in all magnifications Close slit edges to approx 0 5 mm The borders must be in sharp focus here Completely open slit edges and...

Page 11: ...15004 2 standards into account The EN 60601 1 standard must be observed when using different medical and or non medical electrical devices in combination Compliance of the BM 900 slit lamp with the Di...

Page 12: ...sistance to interfe rence Third party devices may only be connected in compliance with the EN 60601 1 standard F 2 Emitted interference standard table 1 The information is based on the requirements of...

Page 13: ...or hospital environment Surge EN 61000 4 5 1 kV for symmetrical voltages 2 kV for asymmetrical voltages 1 kV for symmetrical voltages 2 kV for asymmetrical voltages Mains power quality should be that...

Page 14: ...ded separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interf...

Page 15: ...minal output of the transmitter W Safe distance according to transmission frequency m 150 kHz 80 MHz D 0 35 80 MHz 800 MHz D 0 7 800 MHz 2 7 GHz D 1 4 0 01 0 1 1 10 100 For transmitters with a nominal...

Page 16: ...Should you have any further questions please contact your HAAG STREIT dealer at http www haag streit com contact contact your distributor html HAAG STREIT AG Gartenstadtstrasse 10 3098 Koeniz Switzerl...

Reviews: