1. Safety

DANGER!

Failure to comply with these instructions may result in material damage 

or pose a danger to patients or users.

WARNING!

These warnings must absolutely be complied with to guarantee 

safe operation of the device and to avoid any danger to users and 

to patients.

 

NOTE!

Important information: please read carefully. 

1.1  Areas of application of the device

The device is intended to use in professional health care facility environment, like

doctor's practices, hospitals and optometrists and opticians premises, except near

of HF surgical equipment and in RF shielded rooms of ME-systems for magnet-

ic resonance imaging. Some portable radio frequency equipment, like cell phones 

or RF telephone equipment including antennas may interference medical devices. 

Such equipment has to be kept in a distance of more than 30 cm (12 inches) from 

any part of the instrument. Inobservance of this precaution may lower the correct 

function of the instrument.

1.2  Ambient conditions

Transport:

Temperature

Air pressure

Relative humidity

from

from

from

−40°C

500 hPa

10%

to

to

to

+70°C

1060 hPa

95%

Storage:

Temperature

Air pressure

Relative humidity

from

from

from

−10°C

700 hPa 

10%

to

to

to

+55°C

1060 hPa

95%

Use:

Temperature

Air pressure

Relative humidity

from

from

from

+10°C

800 hPa 

30%

to

to

to

+35°C

1060 hPa

90%

1.3  Shipment and unpacking

• 

Before you unpack the appliance, check whether the packaging shows traces of 

incorrect handling or damage. If this is the case, notify the transport company that 

has delivered the goods to you. Unpack the equipment together with a represen-

tative of the transport company. Make a report of any damaged parts. This report 

must be signed by you and by the representative of the transport company.

• 

Leave the device in the packaging for a few hours before unpacking it (conden-

sation).

• 

Check the appliance for damage after it is unpacked. Return defective applianc-

es in the appropriate packaging.

• 

Store packaging material carefully so that it can be used for potential returns or 

when moving.

• 

The slit lamp and head rest must be installed on an electrically insulated, fire

-

proof table top.

• 

The rail covers (a) prevent the slit lamp from tilting.

• 

Are the connection parts of the accessories in the correct position (screw con-

nections, quick-release fasteners)?

(a) 

1.4  Installation warnings 

WARNING!

• 

Do not modify this equipment without authorization of the manufacturer.

Installation and repairs may only be performed by trained specialists.

Any third-party device must be connected in compliance with the 

EN 60601-1 standard.

Only original HS replacement parts may be used. 

The device must not be stacked or placed in close proximity to other 

electronic devices.

 


 

1.5  Operation, environment

DANGER!

Never use the device in potentially explosive environments where vola-

tile solvents (alcohol, petrol, etc.) and flammable anaesthetics are in

use.
WARNING!

The device must be switched off after every use. Otherwise there is a

risk of overheating when a protective dust cover is used.

01_IFU_BM900-7200492-04150_eng.indd   4

21.02.2017   09:54:53

Summary of Contents for BM 900

Page 1: ...00 7200492 04150 15 Edition 2017 01 DEUTSCH ENGLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA DOC no 1500 1500 1400209 04000 INSTRUCTIONS FOR USE Slit lamp BM 900 15 Edition 2017 01 01_IF...

Page 2: ...of diseases or trauma which affect the structural properties of the eye Contraindication There is no absolute contraindication for tests with this device Appropriate profes sional judgement and cautio...

Page 3: ...Preparing the patient 7 5 3 Operating the instrument 7 5 4 Setting the filters diaphragms 8 5 5 Fixation star 9 5 6 Microscope and eyepiece 9 6 Decommissioning 9 7 Technical data 9 7 1 Slit illuminat...

Page 4: ...npack the appliance check whether the packaging shows traces of incorrect handling or damage If this is the case notify the transport company that has delivered the goods to you Unpack the equipment t...

Page 5: ...nued use of a product damaged by incorrect handling may lead to personal injury In such a case the manufacturer will not accept any liability Haag Streit does not grant any warranties either expressed...

Page 6: ...Breath shield option 14 Protective cover 15 Illumination unit microscope an gle scale 16 Illumination arm locking screw 17 Microscope arm locking screw 18 Slit width setting screw 19 Slit length diap...

Page 7: ...s and press the rocker switch The green light on the rocker switch illuminates when the device is switched on Turn the rotating knob on the illumination control to a position between 1 and 10 5 Operat...

Page 8: ...d 10 a short mirror 12 is used the illumination turned 90 using the locking screw 20 and tilted in 5 steps us ing the latch 27 and the illumination and microscope turned to the central po sition 0 The...

Page 9: ...Dioptre scale for 16x eyepieces for setting the refraction of the examin er 8 D 46 Dioptre scale for 10x eyepieces for setting the refraction of the examin er 8 D 47 Index white point 48 10x or 25x e...

Page 10: ...from the to the side The structure of the test rod must be in sharp focus in all magnifications Close slit edges to approx 0 5 mm The borders must be in sharp focus here Completely open slit edges and...

Page 11: ...15004 2 standards into account The EN 60601 1 standard must be observed when using different medical and or non medical electrical devices in combination Compliance of the BM 900 slit lamp with the Di...

Page 12: ...sistance to interfe rence Third party devices may only be connected in compliance with the EN 60601 1 standard F 2 Emitted interference standard table 1 The information is based on the requirements of...

Page 13: ...or hospital environment Surge EN 61000 4 5 1 kV for symmetrical voltages 2 kV for asymmetrical voltages 1 kV for symmetrical voltages 2 kV for asymmetrical voltages Mains power quality should be that...

Page 14: ...ded separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interf...

Page 15: ...minal output of the transmitter W Safe distance according to transmission frequency m 150 kHz 80 MHz D 0 35 80 MHz 800 MHz D 0 7 800 MHz 2 7 GHz D 1 4 0 01 0 1 1 10 100 For transmitters with a nominal...

Page 16: ...Should you have any further questions please contact your HAAG STREIT dealer at http www haag streit com contact contact your distributor html HAAG STREIT AG Gartenstadtstrasse 10 3098 Koeniz Switzerl...