3.  Appliance assembly / installation

WARNING!

 

Do not modify this equipment without authorization of the manufacturer.

Installation and repairs may only be performed by trained specialists. 

Contact your HAAG-STREIT representative for installation, repairs and 

modification work on the system. The contact details are available at 

www.haag-streit.com.

Only original HS replacement parts may be used.

 

 

3.1  Microscope and illumination

The slit lamp is packaged and shipped fully assembled. The transport safety devic-

es must be removed before commissioning.

3.2  Power supply

NOTE!

Observe the respective HAAG-STREIT instructions for use. (For further 

information, please contact your HAAG-STREIT dealer).

This device must only be operated with PS-LED and PS-LED HSM 901 

HS power supplies and the RM02 release module.


3.3  Regulating the clearance of the slit width control

The small screw in the centre of the right control knob 

(18) allows you to regulate the friction of the turning 

movement of these adjusting knobs. Turning it slightly to 

the right (in) makes it harder, turning it left (out) makes it 

easier. It should at least be set so hard that the slit can-

not close on its own.

18

4. Commissioning

The device can be switched on and off using the mains power switch on the device 

power supply. The green light on the rocker switch illuminates when the device is 

switched on.

4.1  Switching on the device

• 

Connect the power supply up to the mains and press the rocker switch. The 

green light on the rocker switch illuminates when the device is switched on.

• 

Turn the rotating knob on the illumination control to a position between '1' and 

'10'.

5. Operation

5.1  Setting the eyepieces

NOTE!

The eyepieces must be individually set prior to the first examination in 

accordance with the refraction of the examiner. Insert the provided fo-

cus test rod (32) in place of the protective cover (14) and turn its black 

projection surface at a right angle to the microscope axis. Return the 

illumination and microscope to the central position (0°).

14. Protective cover

32. Test rod

33. Knurled ocular refraction ring with 

dioptre scale

32

14

33

• 

Each eyepiece should be set individually by turning the knurled ocular refraction 

ring with dioptre scale (33) such that the projected slit can be seen in focus. The 

setting is performed from the (+) to the (−) side at low magnification.

• 

The pupil distance should then be set on the microscope.

5.2  Preparing the patient

• 

In order to attain a solid basis for the forehead and chin to rest on, the table 

height should be selected such that the patient sits bent over forward.

• 

To ensure that only the part of the eye being examined is illuminated, the slit 

height should be set accordingly.

• 

Parts which come into contact with the patient should be cleaned with a dry cloth 

prior to every use.

• 

The lamp must be switched off after every examination.

5.3  Operating the instrument

WARNING!

The device must be switched off after every use. Otherwise there is a 

risk of overheating when a protective dust cover is employed.

• 

Use the turn screw (7) to set the chin rest (6) in such a way that the patient’s 

eyes are at the same height as the black mark (4) on the sides of the head rest. 

01_IFU_BM900-7200492-04150_eng.indd   7

21.02.2017   09:54:56

Summary of Contents for BM 900

Page 1: ...00 7200492 04150 15 Edition 2017 01 DEUTSCH ENGLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA DOC no 1500 1500 1400209 04000 INSTRUCTIONS FOR USE Slit lamp BM 900 15 Edition 2017 01 01_IF...

Page 2: ...of diseases or trauma which affect the structural properties of the eye Contraindication There is no absolute contraindication for tests with this device Appropriate profes sional judgement and cautio...

Page 3: ...Preparing the patient 7 5 3 Operating the instrument 7 5 4 Setting the filters diaphragms 8 5 5 Fixation star 9 5 6 Microscope and eyepiece 9 6 Decommissioning 9 7 Technical data 9 7 1 Slit illuminat...

Page 4: ...npack the appliance check whether the packaging shows traces of incorrect handling or damage If this is the case notify the transport company that has delivered the goods to you Unpack the equipment t...

Page 5: ...nued use of a product damaged by incorrect handling may lead to personal injury In such a case the manufacturer will not accept any liability Haag Streit does not grant any warranties either expressed...

Page 6: ...Breath shield option 14 Protective cover 15 Illumination unit microscope an gle scale 16 Illumination arm locking screw 17 Microscope arm locking screw 18 Slit width setting screw 19 Slit length diap...

Page 7: ...s and press the rocker switch The green light on the rocker switch illuminates when the device is switched on Turn the rotating knob on the illumination control to a position between 1 and 10 5 Operat...

Page 8: ...d 10 a short mirror 12 is used the illumination turned 90 using the locking screw 20 and tilted in 5 steps us ing the latch 27 and the illumination and microscope turned to the central po sition 0 The...

Page 9: ...Dioptre scale for 16x eyepieces for setting the refraction of the examin er 8 D 46 Dioptre scale for 10x eyepieces for setting the refraction of the examin er 8 D 47 Index white point 48 10x or 25x e...

Page 10: ...from the to the side The structure of the test rod must be in sharp focus in all magnifications Close slit edges to approx 0 5 mm The borders must be in sharp focus here Completely open slit edges and...

Page 11: ...15004 2 standards into account The EN 60601 1 standard must be observed when using different medical and or non medical electrical devices in combination Compliance of the BM 900 slit lamp with the Di...

Page 12: ...sistance to interfe rence Third party devices may only be connected in compliance with the EN 60601 1 standard F 2 Emitted interference standard table 1 The information is based on the requirements of...

Page 13: ...or hospital environment Surge EN 61000 4 5 1 kV for symmetrical voltages 2 kV for asymmetrical voltages 1 kV for symmetrical voltages 2 kV for asymmetrical voltages Mains power quality should be that...

Page 14: ...ded separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interf...

Page 15: ...minal output of the transmitter W Safe distance according to transmission frequency m 150 kHz 80 MHz D 0 35 80 MHz 800 MHz D 0 7 800 MHz 2 7 GHz D 1 4 0 01 0 1 1 10 100 For transmitters with a nominal...

Page 16: ...Should you have any further questions please contact your HAAG STREIT dealer at http www haag streit com contact contact your distributor html HAAG STREIT AG Gartenstadtstrasse 10 3098 Koeniz Switzerl...