Stereo-microscope
Stereo angle
13°
Magnification changer
1x and 1.6x
Magnification
with ocular 10x (standard)
10x and 16x
with ocular 25x
25x and 40x
Range of adjusting eye-pieces
+ 6 to -6 diopters
Inter pupillary distance
54 - 94 mm
Interface video camera
C-mount
Instrument base
Operation Single
handed
3-dimensional
opera-
tion of the guide lever
Adjustment of the instrument base 80 mm (length)
30 mm (height)
100 mm (side)
Electric supply
Class II, type B
230 V~, 0.13 - 0.21 A
or 115 V~, 0.25 - 0.42 A
50 - 60 Hz
Miscellaneous
Weight
8.1 kg (without
power supply, headrest
and
options)
Classification
Against electric shock
Class I, type B
Against water
IPX0
Sterilisation and disinfection
• for the slit lamp
not necessary
• for the headrest
use chin paper and
clean the forehead
with cotton soaked
in
alcohol
Operation mode
continuous
Power supply
Input
AC 115V, ±10% / 0.25 - 0.42A
AC 230V, ±10% / 0.13 - 0.21A
50 - 60Hz
Output 1
DC 7 - 14.4V / max. 36W
Output 2
DC 5V / 1W
Classification
• Class II
•
IPX0
• continuous operation
•
equipment
not suitable
for use
in the presence of a flammable
anaesthetic mixture