Unit Description
8
5.2
Safety class 2 according to DIN 12879
A variably adjustable excess temperature protection
and independent low liquid level protection which is
preset to the lowest level allow the usage of different
heat transfer liquids.
If a safety feature is triggered
…
•
the Fault Identification System (FIS) and an acoustic
signal indicate the fault,
•
the safety-relevant components of the heating unit
(heating element and motor) are switched off immedi-
ately i.e. the safety circuit transfers the unit to a stable,
safe condition,
•
the heat transfer liquid in the heating unit gradually ad-
justs to ambient temperature, but
…
For units with switched on compressor
cooling, this cooling remains functional
and thus cools the heat transfer liquid to the
lowest reachable temperature.
5.3
Applications
The circulator is intended for
…
•
temperature controlling samples within the circulator’s
own bath or for
…
•
temperature controlling closed temperature control cir-
cuits such reactors, heat exchangers or similar objects.
Open vessels cannot be temperature controlled as these cir-
culators are only equipped with a pressure pump.