The following table lists sensor warning messages, where #n is defined as follows:

#0 refers to the CO

2

 sensor

#1 refers to the Brix/Diet sensor

#2 refers to the O

2

 sensor

The last three messages in the table apply to the Brix/Diet sensor only.

Sensor Warning Messages

Message on 

Instrument LCD

LCD 

Code

PC 

Event 

List

High/

Low 

Relay

Hi-Hi/

Lo-Lo 

Relay

Reason

Action

#n CHECK 

THE SENSOR

Out

Sensor 

out

On

On

Sensor disconnected

Connect sensor

#n WARNING 

THERMAL 

CUTOFF

Hot

Sensor 

too hot

On

On

Sensor Thermal cutoff

Check process

#0 SENSOR 

PURGE 

FAILURE

Prg

Overflow On

On

At the beginning of a 

measurement cycle if 

V3beg< -2500mV or 

V3beg> +4450mV

Restore purge gas

#0 RAMP LIMIT 

VIOLATED

Lmt

Overflow On

On

At the end of a 

measurement cycle if 

V3end< -2500mV or 

V3end> +4450mV

Check membrane

#0 PURGE 

VOLTAGE 

DRIFTING

Chk

N/A

Off

Off

At the beginning of a 

measurement cycle if 

V3beg< -2000mV or 

V3beg> +3950mV OR 

at the end of a mea-

surement cycle if 

V3end< -2000mV or 

V3end> +3950mV

Check purge gas

#n ATTENTION 

LOW LIMIT

LoL

Low limit On

Off

Low limit alarm

Check process

#n ATTENTION 

HIGH LIMIT

HiL

High 

Limit

On

Off

High limit alarm

Check process

#n ATTENTION 

LOWLOW LIMIT

LLL

Lo/Lo 

limit

On

On

LowLow limit alarm

Check process

#n ATTENTION 

HIGHHIGH LIMIT HHL

Hi/Hi 

limit

On

On

HighHigh limit alarm

Check process

OSCILLATOR 

STOPPED

Osc

Overflow On

On

Brix/Diet sensor 

malfunction

Check sensor

connections

U-TUBE EMPTY

Emp

Overflow On

On

Brix/Diet sensor 

U-tube empty

Restore sample

DEGASSING IN 

U-TUBE

Deg

Overflow On

On

Sample drop-off in 

U-tube

Restore sample

Summary of Contents for ORBISPHERE 3624

Page 1: ...Operator Quick Guide ORBISPHERE 3624 Revision F 03 10 2008...

Page 2: ...accidental modifications The Display has a two line liquid crystal display LCD with 16 characters per line A Light Button when pressed provides illumination to the LCD for three minutes Four Function...

Page 3: ...h active LINE which are activated via the Start Line A B command in the Configuration menu or by the A B buttons on the toolbar at the top left of the screen The Numerical Indicator Panel shows inform...

Page 4: ...utput From the General window these include the title bars color or black and white the data displayed in either or both tabular and graphical form Choose the styles that suit you best Select the Plot...

Page 5: ...ts the instrument sets an alarm SERIAL OUTPUT instrument only The instrument serial output is normally used to interface to the PC for analysis of measurement data via the ProBrix Plus PC Program CONT...

Page 6: ...roducts on both Line A and B Autosave This option allows you to select the time delay for automatically saving measurement data Set to None no Autosave required End of Run save only on completion or a...

Page 7: ...ts three calibration cycles The gas and temperature measurements are displayed along with the current calibration cycle number On completion of the three cycles the instrument then shows the sensitivi...

Page 8: ...e diet and T A products only Before doing a calibration it is necessary to check your water calibration coefficient first This can be done during the next rinse down period Make sure you have set the...

Page 9: ...ere agent System does not start No display or backlight Improper line voltage Check instrument voltage Select proper voltage 115V or 230V on power supply card Blown fuse Unplug instrument and replace...

Page 10: ...bble should evolve during the measurement ramp If solenoid leaks more than 1 bubble in 15 seconds return sensor to Orbisphere Grounding mismatch between Orbisphere instrument and electronic accessorie...

Page 11: ...VIOLATED Lmt Overflow On On At the end of a measurement cycle if V3end 2500mV or V3end 4450mV Check membrane 0 PURGE VOLTAGE DRIFTING Chk N A Off Off At the beginning of a measurement cycle if V3beg...

Page 12: ......

Reviews: