Page 3

Switching on the indicating instrument first shows
the software version and date on the LCD, and
then switches to a 

Configuration Display

showing the gas being measured, the purge gas
required for the CO

2

 sensor, and the membrane

model number for this sensor. 

A configuration
screen appears
briefly, before
switching to a

C a l i b r a t i o n

Coefficients

 screen for several seconds before

finally switching to an 

O2 Measurement Display

.

This display then alternates every few seconds

with one of the CO

2

 or °Brix/%Diet measurement

displays (illustrated left).
Press the 

×

  

Ø

 keys to switch between these

t h r e e   m e a s u r e m e n t   d i s p l a y s .   Y o u   m a y
experience a delay before the display changes,
depending on the purge/measurement cycle of
the CO

2

 sensor. Press the 

ESC

 key to exit from

these displays and move to the 

Main Menu

.

The O

2

 and Brix/Diet LED displays are driven by

analog outputs from the instrument. These
displays are updated at regular time intervals but
may vary slightly from the instrument LCD's
displayed values.

Measurement - Instrument

Measurement - PC

Once the ProBrix Plus PC Program software is installed on your PC, it is recommended that no screen
savers are set and that the PC is not switched off unless absolutely necessary. 
The PC Program maintains a window for each active LINE, which are activated via the 

Start Line A/B

command in the 

Configuration menu

, or by the 

A/B buttons

 on the toolbar at the top left of the screen.

The 

Numerical Indicator Panel

shows information relating to the last
measurement, and is divided into
four, containing the CO

2

 data, the

Brix/Diet data, the O

2

 data and the

temperature information. For each of
these, the units of measurement, the
most recent measurement value and
a vertical bar graph are shown.
The background colors show: 

GREEN

 measurements are within 

High and Low limits

ORANGE

 measurements are 

outside High or Low limits, but inside 
High/High and Low/Low limits

RED

 measurements are outside 

High/High or Low/Low limits or an 
error has occurred

The 

Status Panel

 is displayed below the indicator panel and shows the status for CO

2

, Brix/Diet and O

2

indicating 

Standby

Measure

 or 

Fault

.

The 

Graphic Panel

 contains four charts showing the measurements over a one-hour period. Any one-

hour period within the last 7½ hours can be viewed using the horizontal scroll bar. The lines on the graph
are color-coded as per the description at the top of the panel. Placing the mouse cursor over a data point
on the graph will cause the cursor to change to a hand symbol. Clicking the mouse when the hand
symbol is visible will cause a message box to appear giving the values of the measurements, including
the time they were taken.
The 

Buttons Panel

 at the bottom of the window contains eight additional options.

Summary of Contents for ORBISPHERE 3624

Page 1: ...Operator Quick Guide ORBISPHERE 3624 Revision F 03 10 2008...

Page 2: ...accidental modifications The Display has a two line liquid crystal display LCD with 16 characters per line A Light Button when pressed provides illumination to the LCD for three minutes Four Function...

Page 3: ...h active LINE which are activated via the Start Line A B command in the Configuration menu or by the A B buttons on the toolbar at the top left of the screen The Numerical Indicator Panel shows inform...

Page 4: ...utput From the General window these include the title bars color or black and white the data displayed in either or both tabular and graphical form Choose the styles that suit you best Select the Plot...

Page 5: ...ts the instrument sets an alarm SERIAL OUTPUT instrument only The instrument serial output is normally used to interface to the PC for analysis of measurement data via the ProBrix Plus PC Program CONT...

Page 6: ...roducts on both Line A and B Autosave This option allows you to select the time delay for automatically saving measurement data Set to None no Autosave required End of Run save only on completion or a...

Page 7: ...ts three calibration cycles The gas and temperature measurements are displayed along with the current calibration cycle number On completion of the three cycles the instrument then shows the sensitivi...

Page 8: ...e diet and T A products only Before doing a calibration it is necessary to check your water calibration coefficient first This can be done during the next rinse down period Make sure you have set the...

Page 9: ...ere agent System does not start No display or backlight Improper line voltage Check instrument voltage Select proper voltage 115V or 230V on power supply card Blown fuse Unplug instrument and replace...

Page 10: ...bble should evolve during the measurement ramp If solenoid leaks more than 1 bubble in 15 seconds return sensor to Orbisphere Grounding mismatch between Orbisphere instrument and electronic accessorie...

Page 11: ...VIOLATED Lmt Overflow On On At the end of a measurement cycle if V3end 2500mV or V3end 4450mV Check membrane 0 PURGE VOLTAGE DRIFTING Chk N A Off Off At the beginning of a measurement cycle if V3beg...