background image

Page 4

File Menu

This menu contains a standard list of File options
to allow you to open a file of historical data. When
selecting this menu, the four most recently
accessed files are listed. 

Open

ing a file will display the initial screen

showing data in either tabular or graphical format,
or a combination of the two. You can resize these
areas by dragging the slider bar left or right with
the mouse, or alternatively using the 

Split

command in the 

View Menu

.

Any changes you make can then be 

Save

d from

this menu, as well as being saved to a different file
using the 

Save As

 option. You can also use the

Print Setup

 option to set up the printer you wish

to use during the process for any hard-copy
output. The 

Properties

 option allows you to view

standard information about the file and product.

Table Menu

While in the Table Window, you can right click the
mouse for quick access to some of the Table
Menu options.
A copy option lets you select table contents as a
text file and copy them to the clipboard for pasting
into other applications (e.g. Excel). The search
option allows you to search for a specified alarm
or event chosen from a drop-down list.
You can change what measurements are
displayed in the table, the formatting for the

numerical values and color options for highlighted
rows. Also, you can adjust the default and sizes of
selected rows or columns, and define the
measurements you want to view.

Graph Menu

T h i s   o p t i o n   g i v e s   y o u   m u l t i p l e   w a y s   o f
customizing the output.
From the 

General

 window, these include the title

bars, color or black and white, the data displayed
in either or both tabular and graphical form.
Choose the styles that suit you best. Select the

Plot Style

 tab to define how the graphs should be

displayed (bar graph, line graph, etc.). These can
be different for each measurement.
The 

Subsets

 tab allows you to define which sets

of measurement data you wish to display. Select
the 

Points

Font

 and 

Color

 tabs and personalize

your graphical displays to suit your requirements.

Statistics

The statistics option will show the measurements
in graphical and/or tabular format depending on
how you customize the output.
Double click anywhere in the window to display
the customize menu which allows you to
customize your display. Right click the mouse to
display a list of quick access display options.
Click the hand symbol (when available) for a pop-
up display of statistics applicable at that point.

PBViewer Program

The 

PBViewer Plus

 program allows

you to view statistical information that
has been saved by the PC Program
from the current or previous product
runs, in a graphical and tabular
format. You can start PBViewer Plus
from within the ProBrix Plus PC
Program by selecting 

PBViewerPlus

from the 

Tools

 menu. This will

display details for the current Product
Run. 
The program can also be run on a
stand-alone basis and can be started
by double-clicking the Desktop icon.
Starting the program this way will

display an empty details screen, until you select a file to open from the 

File Menu

.

There are three standard views available - the Table Window, the Graph Window or a combination of
both. The illustration shows the combined view.

Summary of Contents for ORBISPHERE 3624

Page 1: ...Operator Quick Guide ORBISPHERE 3624 Revision F 03 10 2008...

Page 2: ...accidental modifications The Display has a two line liquid crystal display LCD with 16 characters per line A Light Button when pressed provides illumination to the LCD for three minutes Four Function...

Page 3: ...h active LINE which are activated via the Start Line A B command in the Configuration menu or by the A B buttons on the toolbar at the top left of the screen The Numerical Indicator Panel shows inform...

Page 4: ...utput From the General window these include the title bars color or black and white the data displayed in either or both tabular and graphical form Choose the styles that suit you best Select the Plot...

Page 5: ...ts the instrument sets an alarm SERIAL OUTPUT instrument only The instrument serial output is normally used to interface to the PC for analysis of measurement data via the ProBrix Plus PC Program CONT...

Page 6: ...roducts on both Line A and B Autosave This option allows you to select the time delay for automatically saving measurement data Set to None no Autosave required End of Run save only on completion or a...

Page 7: ...ts three calibration cycles The gas and temperature measurements are displayed along with the current calibration cycle number On completion of the three cycles the instrument then shows the sensitivi...

Page 8: ...e diet and T A products only Before doing a calibration it is necessary to check your water calibration coefficient first This can be done during the next rinse down period Make sure you have set the...

Page 9: ...ere agent System does not start No display or backlight Improper line voltage Check instrument voltage Select proper voltage 115V or 230V on power supply card Blown fuse Unplug instrument and replace...

Page 10: ...bble should evolve during the measurement ramp If solenoid leaks more than 1 bubble in 15 seconds return sensor to Orbisphere Grounding mismatch between Orbisphere instrument and electronic accessorie...

Page 11: ...VIOLATED Lmt Overflow On On At the end of a measurement cycle if V3end 2500mV or V3end 4450mV Check membrane 0 PURGE VOLTAGE DRIFTING Chk N A Off Off At the beginning of a measurement cycle if V3beg...