Page 9
The following table lists possible measurement errors, which could be caused by poor handling of the
instrument and/or identifiable installation conditions.
Measurement Troubleshooting
Symptom
Probable Cause
Corrective Action
Expected Result
Message
Out
Sensor not connected
Check sensor cable
connection (at sensor and
instrument end)
Measurement should return after
approximately 20 seconds
Message
Prg
(CO
2
)
Purge pressure too low or
wrong purge gas
Restore correct purge pres-
sure and gas (refer to the
TC
Sensors - Maintenance &
Installation Manua
l for
details). Push the small plas-
tic tube firmly into the purge
exit and check purge flow rate
(over 3 bubbles per second.)
Check voltage in continuous
purge mode.
In continuous purge mode,
voltage should be in the range
-2.5V to +4.5V. If purge flow rate
is too small, or if voltage remains
above 4.5V, contact your
Orbisphere agent
Damaged sensing
element caused by
mishandling during
membrane replacement
For experienced
service
personnel only
: Remove
membrane and support.
Check for broken wires of the
thermal conductivity chip
under microscope
Contact your Orbisphere agent
System does not
start: No display
or backlight
Improper line voltage
Check instrument voltage
Select proper voltage (115V or
230V) on power supply card
Blown fuse
Unplug instrument and
replace blown fuse.
Instrument should start up. If not,
or if fuse blows again, contact
your Orbisphere agent
High sample
temperature
Blockage of sample flow
causing heat buildup from
pump
Check outlet of sample flow
for blockage. If a check valve
is in use, verify proper
orientation
Temperature should decrease as
normal sample flow resumes
Voltage drift
during
measurement
(CO
2
)
Purge gas pressure too
low
Set 2 bar gauge pressure or
adjust pressure in order to get
over 3 bubbles per second
from exit tube
Measurement will return to
expected value
Low reading
(CO
2
)
Leak in purge inlet line
Check line for leaks with
soapy water
Expected reading
High reading
(CO
2
)
Erroneous calibration in
gas with a wet membrane
protection grill
Recalibrate in liquid, or,
dry out the membrane and
calibrate in gas
Expected reading
Totally erratic
readings (CO
2
)
Water condensation into
the thermal conductivity
element due to purge
failure or humid purge gas
Blow
dry
purge gas in "contin-
uous purge mode" overnight.
Resume measurement mode.
In most cases, sensor operates
properly. But condensation may
result in serious damage,
requiring service