3624 ProBrix Plus - Installation
Operator Manual
ORBISPHERE
1.4 Connections
1.4.1 Sampling Module Connections
• Connect the two filling line connectors to the sample inlet and outlet connections
on the side of the sampling module. Use ½" inside diameter, rigid stainless-steel
tubing or (optionally) flexible braided stainless-steel tubing. Connect this tubing to
isolation valves that are welded to the filling line.
• The tubing should be as straight as possible with no up or down bends, in order to
prevent any degassing. If the system is isolated from the filling line by valves, wait
for a Cleaning in Place (CiP) cycle before opening them.
• Ensure that the liquid detector is installed in the inlet line and that its cable is
connected to the power unit located inside the sampling module.
• Place the N
2
purge gas cylinder in its fixture on the right of the sampling module,
holding it in place with the straps provided, and connect the gas tubing to the inlet
above the straps. Check the pressure on the pressure regulator of the cylinder (3
to 5 bar gauge) and adjust the pressure on the internal purge gas pressure
regulator (inside the sampling module) to 2 bar gauge. Check all gas fittings for
leaks with soapy water (or a leak detector spray), from the gas cylinder to the CO
2
sensor.
• Connect the serial cable from the PC to the instrument (see
). Connect the hold switch and alarm relays, if used, to their respective
terminals in the indicating instrument (see
“Instrument Connectors” on page 17
).
• Connect the 115 or 230 VAC mains power supply to the power unit.
Fig 1-10:
Sampling Module External Filling
Summary of Contents for ORBISPHERE 3624
Page 1: ...Operator Manual ORBISPHERE 3624 Revision F 03 10 2008 ...
Page 2: ......
Page 14: ...8 of 96 Manual Overview 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 32: ...26 of 96 Installation 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 60: ...54 of 96 PC Program Setup 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 68: ...62 of 96 Calibrations 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 82: ...76 of 96 Accessories and Attachments 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 98: ...92 of 96 Glossary 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 101: ...3624 ProBrix Plus User Notes 95 of 96 Operator Manual ORBISPHERE User Notes ...
Page 102: ......