Installation - 3624 ProBrix Plus
Operator Manual
ORBISPHERE
CAUTION:
You must ground the cable shield securely to the instrument to ensure proper operation.
Do not ground the shield at the PC.
Insert the sub-D 9-pin connector into the communications port at the back of the PC e.g.
COM 3, or COM 4 for a second product line running from the ProBrix Plus PC Program.
Note:
The RS-422/485 card has two ports.
After starting the ProBrix Plus PC Program, click on the
Line A
and/or
Line B
option in
the
Configuration Menu
and verify that the COM port is set to those connected above
(e.g. COM3 and/or COM4) depending on the line(s) you are using (see
).
Once the sampling module and instrument are switched on, communication tests can be
made. Refer to
“Serial Output (instrument only)” on page 40
to enable the serial
communications from the 3624 instrument to the ProBrix Plus PC Program.
Following a product change, a standby message should appear on the ProBrix Plus PC
Program's bar graph within 30 seconds, and a measurement or standby message after
another 30 seconds.
Note:
Do not enable any screen savers on the PC, and for best performance do not shut off
the PC during weekends or shutdowns. The monitor, however, may be switched off at
any time.
Summary of Contents for ORBISPHERE 3624
Page 1: ...Operator Manual ORBISPHERE 3624 Revision F 03 10 2008 ...
Page 2: ......
Page 14: ...8 of 96 Manual Overview 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 32: ...26 of 96 Installation 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 60: ...54 of 96 PC Program Setup 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 68: ...62 of 96 Calibrations 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 82: ...76 of 96 Accessories and Attachments 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 98: ...92 of 96 Glossary 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 101: ...3624 ProBrix Plus User Notes 95 of 96 Operator Manual ORBISPHERE User Notes ...
Page 102: ......