3624 ProBrix Plus - Specifications
Operator Manual
ORBISPHERE
9
Specifications
9.1 General Technical Data
Table 9-1:
General Technical Data
CO
2
Analysis
O
2
Analysis
Brix/Diet
Analysis
Measurement Range
0-15 g/kg, or
0-7 V/V, or
0-1.5%
0.001-80 ppm
0-20° Brix,
0-200% Diet
Accuracy
(CO
2
and O
2
: within ±5°C
of calibration temperature)
±1% of reading, or
±0.012 g/kg, or
±0.006 V/V,
whichever is greater
±1% of reading, or
±1 ppb,
whichever is greater
±0.02° Brix, or
±0.5% Diet
Signal drift
<1% of reading
between service
<1% of reading
between service
Negligible
Startup response time
45 seconds
38 seconds
1 minute (Brix)
3 minutes (Diet)
System refresh rate
15 seconds
Sample flow requirement
Maintained by internal pump
Sample pressure limit
12 bar
Sensor temperature limit
-10°C to +100°C
Temperature compensated
range
0°C to 35°C
Digital output
RS-422/485
Baud rate: 9600, Stop Bit: 1, Start Bit: 0, Parity: None
Analog output
0/4-20 mA
Alarm outputs
2 dry contact relays:
Max voltage: 24 VDC
Max power: 30 W
Life time (24 V, 1 A): 2 million cycles
Hold inputs
Dry contact or 24 VDC depending on configuration
(see
Summary of Contents for ORBISPHERE 3624
Page 1: ...Operator Manual ORBISPHERE 3624 Revision F 03 10 2008 ...
Page 2: ......
Page 14: ...8 of 96 Manual Overview 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 32: ...26 of 96 Installation 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 60: ...54 of 96 PC Program Setup 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 68: ...62 of 96 Calibrations 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 82: ...76 of 96 Accessories and Attachments 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 98: ...92 of 96 Glossary 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Page 101: ...3624 ProBrix Plus User Notes 95 of 96 Operator Manual ORBISPHERE User Notes ...
Page 102: ......