Sonde Setup
Page 15
Section 3
Sonde Setup
The Surveyor 4a can setup the Sonde (Series3 or Series4, only). The items displayed in the
Sonde Setup submenu depends on the Sonde connected to the Surveyor 4a. Some of the choices
and options are not available on some Series 3 Sondes. See the Series 3 Operating Manual for
more information on setup.
Security Level-2 is required to change Sonde Setup features.
1.
Connect the Surveyor 4a to a Sonde.
2.
Press the
OFF/ON
button.
3.
From the Main screen, select
SETUP/CAL
>
SETUP
>
SONDE
. The Sonde Setup submenu will be
displayed.
4.
Highlight the option to be configured. Sections
describes the Sonde Setup submenu
features.
3.1 Selecting Parameters
This feature allows the setup of parameter setup options for the Sonde. The Surveyor 4a will
display the parameters included on the Sonde.
3.2 Turning the Circulator On/Off
Note:
When using a DataSonde 3 or Recorder, this option allows the “stirrer” to be turned on of off.
1.
Highlight Circltr:ON/Off and press
SELECT
.
2.
Choose to turn the circulator on or off using the cursor keys and the
SELECT
key. Select 0
(zero) to disable the circulator or select 1 to enable the circulator.
3.
Press
DONE
when complete.
3.3 Turning the Audio On/Off
Note:
When using a DataSonde 3 or Recorder, this option allows the “Buzzer” to be turned on of off.
1.
Highlight Audio:On/Off and press
SELECT
.
2.
Choose to turn the audio on or off using the cursor keys and the
SELECT
key. Select 0 (zero) to
disable the audio or select 1 to enable the audio.
3.
Press
DONE
when complete.
3.4 Assigning the SDI-12 Address
1.
Highlight SDI:Address and press
SELECT
.
2.
Enter the SDI address using the cursor keys and the
SELECT
key. The SDI address status is
displayed on the bottom right-hand corner of the screen.
3.
Press
DONE
when complete.