21
Section 7
Faults, causes, rectification
7.1 Error messages
Possible sensor errors are displayed by the controller.
7.2 Warnings
Possible warning messages are displayed by the controller.
Table 1: Error messages
Error displayed
Cause
Rectification
*****
No communication with the
controller
Check the connection to the controller
Check the cable to the controller
SENSOR MISSING
FFFFFFFFFFFFF
No communication with the
controller
Check the connection to the controller
Check the cable to the controller
TEMP TOO LOW
Measured temperature < –5 °C
Ensure that the
medium temperature is > –5 °C.
TEMP TOO HIGH
Measured temperature >
+100 °C
Ensure that the
medium temperature is < +100 °C.
COND TOO LOW
Conductivity < 100
µS
/
cm
Ensure that the
conductivity is > 100
µS
/
cm
.
COND TOO HIGH
Conductivity > 500
mS
/
cm
Ensure that the
conductivity is < 500
mS
/
cm
.
RES. TOO LOW
Resistance < 2
Ω
Please contact service.
RES. TOO HIGH
Resistance > 10 k
Ω
Please contact service.
Table 2: Warnings
Error displayed
Cause
Rectification
CAL TOO OLD
The last calibration was more than
180 days ago.
Calibrate the sensor
HUMIDITY BAG
The desiccant bag is more than
1000 days old.
Please contact service.
Summary of Contents for 3798-S sc
Page 1: ...DOC023 52 03252 3798 S sc Digital inductive conductivity sensor User Manual 08 2022 Edition 3...
Page 2: ......
Page 4: ...2 Table of contents...
Page 6: ...4 Technical data...
Page 12: ...10 General safety instructions...
Page 16: ...14 Installation...
Page 22: ...20 Maintenance...
Page 26: ...24 Spare parts...
Page 28: ...26 Warranty and liability...
Page 31: ......