22
Faults, causes, rectification
7.3 Important service data
Data
Minimum
Maximum
CAL DATA
Electrical gain correction
95 %
105 %
Temperature offset correction
– 5 °C
+ 5 °C
Cell constant
2.50
2.00
Signals
Output voltage
Raw measured data
– 1 %
+ 1 %
Counter
Desiccant bag
Operating time
1000 days
MODBUS STATS
Number of communication errors
0
< 1 %
Measurement of fixed
resistance 1 k
Ω
Measured value
990
Ω
1010
Ω